Related Experiment Video
Updated: Jun 27, 2026

ELIME Enzyme Linked Immuno Magnetic Electrochemical Method for Mycotoxin Detection
Published on: October 23, 2009
Enzyme- and label-free cascade isothermal amplification aptasensor for the ultrasensitive detection of ochratoxin A
Ming Li1, Yu Xiao2, Zexuan Wang3
1School of the Environment and Safety Engineering, School of Emergency Management, Jiangsu University, Zhenjiang, 212013, PR China; Jiangsu Collaborative Innovation Center of Technology and Material of Water Treatment, School of Environmental Science and Engineering, Suzhou University of Science and Technology, Suzhou, 215009, PR China.
Background:
Ultrasensitive detection is crucial for the early warning and intervention of risk factors, ultimately benefiting the environment and human health. Low levels of ochratoxin A (OTA) present a hidden yet significant threat, and rapid detection via high-performing biosensors is therefore essential.
Results:
A cascade isothermal amplification aptasensor (CIA-aptasensor) was designed for OTA detection. On the surface of a magnetic bead probe, the OTA level was converted into positively correlated trigger cDNA through its competitive binding with OTA-Apt. The released trigger cDNA activated catalytic hairpin assembly followed by coupling with a hybridization chain reaction to achieve CIA. After adding graphene oxide and SYBR Green I, the background interference was eliminated to specifically obtain OTA-related fluorescence. The ultrasensitive limit of detection was 0.22 pg mL-1, an improvement of 1368-fold over conventional enzyme-linked aptamer sorbent assay by the same OTA-Apt, demonstrating satisfactory reliability and practicability. Thus, the CIA-aptasensor provides an enzyme- and label-free simplified homogeneous system with minimal background interference using isothermal conditions.
Significance:
This study provides a polymerase chain reaction-like approach for enhancing the sensitivity and performance of a biosensor, which could be extended for the application of CIA and label-free signaling strategy to other risk factors.
More Related Videos
03:38Aptamer-Based Target Detection Facilitated by a 3-Stage G-Quadruplex Isothermal Exponential Amplification Reaction
Published on: October 6, 2022
07:59Rapid and Specific Detection of Acinetobacter baumannii Infections Using a Recombinase Polymerase Amplification/Cas12a-based System
Published on: April 25, 2025
Related Concept Videos
Enzyme-Linked Immunosorbent Assay
There are many different types of ELISAs, but they all involve an antibody molecule whose constant region binds an enzyme, leaving the variable region free to bind its specific antigen. Enzyme-substrate reaction allows the antigen to be visualized or quantified.
Transducer Mechanism: Enzyme-Linked Receptors
Major types that are helpful drug targets include: