Related Experiment Video
Updated: Jul 9, 2026

10:10
Development of a Lateral Flow Immunochromatographic Strip for Rapid and Quantitative Detection of Small Molecule Compounds
Published on: November 13, 2021
7.8K
A Highly Sensitive Lateral-Flow Strip Using Latex Microspheres to Detect NGAL in Urine Samples
Paweena Tunakhun1,2, Sawinee Ngernpimai3, Patcharaporn Tippayawat2
1Biomedical Sciences, Graduate School, Khon Kaen University, Khon Kaen 40002, Thailand.
ACS Omega
|September 2, 2024
Summary
A new lateral flow strip (LFS) test using latex microspheres can rapidly detect urinary Neutrophil Gelatinase-Associated Lipocalin (NGAL) for early kidney disease diagnosis at the point-of-care.
Area of Science:
- Biomarker Discovery
- Diagnostic Assay Development
- Point-of-Care Testing
Background:
- Global increase in kidney disease incidence necessitates early detection methods.
- Neutrophil Gelatinase-Associated Lipocalin (NGAL) is a promising biomarker for acute kidney injury (AKI) and chronic kidney disease (CKD).
- Existing diagnostic methods may lack speed and cost-effectiveness for widespread early screening.
Purpose of the Study:
- To develop a rapid, point-of-care lateral flow strip (LFS) for detecting urinary NGAL.
- To utilize latex microspheres (LMs) as a visual label in a lateral flow immunoassay (LFIA).
- To evaluate the performance and stability of the developed NGAL LFS for clinical utility.
Main Methods:
- Development of a lateral flow immunoassay (LFIA) incorporating latex microspheres (LMs) for NGAL detection.
- Testing the LFS for sensitivity, specificity, and interference using urine samples.
- Comparison of LFS results with the established enzyme-linked immunosorbent assay (ELISA) method.
Main Results:
- The developed LMs-LFS detected urinary NGAL within 15 minutes with high specificity.
- Visual limit of detection (vLOD) was 18.75 ng/mL, and the limit of detection (LOD) was 1.65 ng/mL.
- High correlation observed between LMs-LFS and ELISA (R² = 0.973, n=60), with good room-temperature stability (≥6 months).
Conclusions:
- The developed latex microsphere-based lateral flow strip (LMs-LFS) is a rapid and sensitive tool for urinary NGAL detection.
- This assay demonstrates potential for point-of-care diagnosis and monitoring of kidney disorders.
- The LMs-LFS offers a stable, specific, and cost-effective alternative for renal health assessment.