Related Experiment Video
Updated: Jun 14, 2025

Fabrication of Carbon Nanotube High-Frequency Nanoelectronic Biosensor for Sensing in High Ionic Strength Solutions
Published on: July 22, 2013
Hive-shaped carbon nanotubes coated with MoSe2as an electrochemical sensor for cholesterol
Lei Hu1, Yu-Chiao Huang1, Yung-Jui Huang1
1Department of Electronic and Computer Engineering, National Taiwan University of Science and Technology, Taipei 106335, Taiwan.
Abstract:
The study utilized transition metal chalcogenide, molybdenum diselenide (MoSe2), for application in the field of bioelectrochemical sensing. The MoSe2was combined with carbon nanotubes (CNTs) by chemical vapor deposition to enhance the specific surface area and improve the detection sensitivity. To further increase the contact area between the electrolyte and the electrode, photolithography techniques were employed to fabricate hive-shaped CNTs, thereby enhancing the specific surface area. Next, cholesterol oxidase (ChOx) was coated onto the electrode material, creating a cholesterol biosensor. Cyclic voltammetry was utilized to detect the concentration of cholesterol. The experiment involved segmented testing for cholesterol concentrations ranging from 0μM to 10 mM. Excellent sensitivity, low detection limits, and high accuracy were achieved. In the cholesterol concentration range of 0μM-100μM, the experiment achieved the highest sensitivity of 4.44μAμM⋅cm-2. Consequently, all data indicated that ChOx/MoSe2/CNTs functioned as an excellent cholesterol sensor in the study.