Monte Carlo Model Validation of 6MV Beam of OMID, the First Iranian Linear Accelerator
Mohammad Amin Abdoli1, Maryam Hassanvand1, Navid Nejatbakhsh2
1Department of Physics, Isfahan University of Technology, Isfahan, Iran.
Abstract:
Monte Carlo (MC) techniques are regarded as an accurate method to simulate the dose calculation in radiotherapy for many years. The present paper aims to validate the simulated model of the 6-MV beam of OMID linear accelerator (BEHYAAR Company) by EGSnrc codes system and also investigate the effects of initial electron beam parameters (energy, radial full width at half maximum, and mean angular spread) on dose distributions. For this purpose, the comparison between the calculated and measured percentage depth dose (PDD) and lateral dose profiles was done by gamma index (GI) with 1%-1 mm acceptance criteria. MC model validating was done for 3 cm × 3 cm, 5 cm × 5 cm, 8 cm × 8 cm, 10 cm × 10 cm, and 20 cm × 20 cm field sizes. To study the sensitivity of model to beam parameters, the field size was selected as 10 cm × 10 cm and 30 cm × 30 cm. All lateral dose profiles were obtained at 10 cm. Excellent agreement was achieved with a 99.2% GI passing percentage for PDD curves and at least 93.8% GI for lateral dose profiles for investigated field sizes. Our investigation confirmed that the lateral dose profile severely depends on the considered source parameters in this study. PDD only considerably depends on the initial electron beam energy. Therefore, source parameters should not be specified independently. These results indicate that the current model of OMID 6-MV Linac is well established, and the accuracy of the simulation is high enough to be used in various applications.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
08:31Sample Preparation and Experimental Design for In Situ Multi-Beam Transmission Electron Microscopy Irradiation Experiments
Published on: June 27, 2022
Related Concept Videos
Beams with Symmetric Loadings
The M/EI...
Beams with Unsymmetric Loadings
The first moment-area theorem determines the slope at any point on the beam. This theorem indicates that the change in slope between two points on a beam...
Atomic Absorption Spectroscopy: Radiation and Light Sources
Two common narrow-range 'line' sources used in AAS are hollow-cathode lamps (HCLs) and...
Impact Loading on a Cantilever Beam
When an object is dropped onto the free end of a cantilever, its potential energy due to gravity is...
Deformation of a Beam under Transverse Loading
The insights from the bending moment diagram extend to...
Atomic Emission Spectroscopy: Instrumentation