Focal Pulsed Field Ablation for Premature Ventricular Contractions: A Multicenter Experience
Domenico Giovanni Della Rocca1,2, María Cespón-Fernández1,3, Ahmad Keelani4
1Heart Rhythm Management Center, Postgraduate Program in Cardiac Electrophysiology and Pacing, Universitair Ziekenhuis Brussel-Vrije Universiteit Brussel, European Reference Networks Guard-Heart, Belgium (D.G.D.R., M.C.-F., L.P., A.A., E.S., G. Bala, J.S., G.V., A. Sorgente, C.A., I.O., A.D.M., G.-B.C., C.d.A., A. Sarkozy).
Background:
Pulsed field ablation (PFA) is a novel technology for catheter-based atrial arrhythmia treatment. Evidence of its application for ventricular arrhythmia ablation is still limited. In this study, we describe the feasibility and efficacy of focal PFA for premature ventricular contraction (PVC) ablation.
Methods:
A prospective cohort of 20 patients referred for PVC ablation at 2 centers was enrolled, regardless of the presence of structural heart disease, PVC morphology, or previous ablation attempts. All procedures were performed using the CENTAURI System in combination with contact force sensing catheters and 3-dimensional electroanatomical mapping systems. Energy output and the number of applications were left to the operator's discretion.
Results:
Eleven (55%) procedures were conducted under general anesthesia, 6 (30%) under deep sedation, and 3 (15%) under light sedation. Muscular contraction was observed in one case (5%). Median procedural and fluoroscopy times were 95.5 and 6.55 minutes, respectively. The median number of PFA applications was 8 with a median contact force of 10g. A statistically significant (76%) reduction was observed in mean peak-to-peak bipolar electrogram voltage before and after ablation (0.707 versus 0.098 mV; P=0.008). Ventricular irritative firing was observed in 11 (55%) patients after PFA. The median follow-up was 120 days. Acute procedural success was achieved in 17 of 20 (85% [95% CI, 0.70-1]) patients. Two of the patients with procedural failure had late success with >80% clinical PVC burden suppression during follow-up, and 2 of 17 patients with acute success had late PVC recurrence, which accounts for a total of 17 of 20 (85% [95% CI, 0.70-1]) patients with chronic success. Transient ST-segment depression occurred in 1 patient, and the right bundle branch block was induced in 2 others (permanently only in one case).
Conclusions:
PVC ablation using a focal PFA is feasible, effective, and safe, with promising acute and long-term results in several ventricular locations. Irritative firing is frequently observed. Coronary evaluation should be considered when targeting the outflow tract.
More Related Videos
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
06:57Ablation of Ischemic Ventricular Tachycardia Using a Multipolar Catheter and 3-dimensional Mapping System for High-density Electro-anatomical Reconstruction
Published on: January 31, 2019
Related Concept Videos
Cardiopulmonary Resuscitation III: AED Use
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias