Performance of MALDI-TOF MS, real-time PCR, antigen detection, and automated biochemical testing for the

Stuart Campbell1, Brooke Taylor2, Dimitrios Menouhos2

  • 1Department of Infectious Diseases, Division of Medicine, Royal Darwin Hospital, Darwin, Australia.

PubMed

Insights

Matrix-assisted laser desorption/ionization time of flight mass spectrometry (MALDI-TOF MS) accurately identifies Burkholderia pseudomallei, the cause of melioidosis. This method is now integrated into laboratory workflows for reliable B. pseudomallei detection.

Area of Science:

  • Clinical Microbiology
  • Infectious Diseases
  • Diagnostic Technology

Background:

  • Burkholderia pseudomallei is the causative agent of melioidosis, a serious disease endemic to Southeast Asia and Australia, with expanding geographical reach.
  • Accurate and timely laboratory identification of B. pseudomallei is crucial for patient management, yet can be challenging, leading to potential misdiagnoses.
  • Existing diagnostic methods like automated biochemical testing and MALDI-TOF MS have reported identification inaccuracies.

Purpose of the Study:

  • To validate the performance of an updated MALDI-TOF MS database for identifying Burkholderia pseudomallei.
  • To compare the accuracy of MALDI-TOF MS against automated biochemical testing (Vitek 2 GN card), quantitative real-time PCR (qPCR), and lateral flow immunoassay (LFA).
  • To assess the utility of MALDI-TOF MS for differentiating B. pseudomallei from closely related Burkholderia species.

Main Methods:

  • A derivation cohort of geographically diverse B. pseudomallei and related Burkholderia species was tested.
  • A prospective validation cohort of clinical isolates, including B. pseudomallei and B. cepacia complex, was analyzed.
  • Identification performance was evaluated using MALDI-TOF MS with a 99.9% certainty cutoff, Vitek 2 GN card, qPCR targeting the type III secretion system, and LFA for capsular polysaccharide antigen.

Main Results:

  • MALDI-TOF MS demonstrated a sensitivity of 1.0 and specificity of 1.0 for B. pseudomallei identification and differentiation when using a 99.9% certainty threshold.
  • Automated biochemical testing showed lower performance with a sensitivity of 0.83 and specificity of 0.88.
  • Both qPCR and LFA achieved perfect identification of all B. pseudomallei isolates without any false positives.

Conclusions:

  • The updated MALDI-TOF MS database provides highly accurate identification and differentiation of Burkholderia pseudomallei.
  • MALDI-TOF MS is a reliable, fast, and cost-effective method for B. pseudomallei identification, surpassing automated biochemical testing.
  • The laboratory has incorporated MALDI-TOF MS into its routine workflow for B. pseudomallei identification due to its high accuracy.