A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
Choose Your Collimator Wisely: Inappropriate Collimator Selection During a 177Lu-DOTATATE Posttreatment Scan
Justin G Peacock1,2, Taylor S Young3
1Department of Radiology and the Radiological Sciences, Uniformed Services University, Bethesda, Maryland; justin.peacock@usuhs.edu.
Abstract:
Proper collimator selection is critical to obtaining high-quality, interpretable nuclear medicine images. Collimators help eliminate scatter, which leads to poor spatial resolution and blurry images. We present the case of a posttherapy 177Lu-DOTATATE (Lutathera) patient who was initially imaged with a low-energy, high-resolution collimator routinely used in 99mTc imaging. On image review, the patient was reimaged with the appropriate medium-energy, high-resolution collimator, which resulted in improved image quality. When reviewing the quality of images, it is important to understand modifications to the imaging that can significantly improve image quality and interpretation.