High-resolution quadrupole improves spectral purity and reduces interference from non-target ions in isobaric
Shen Zhang1, J C Yves Le Blanc2, Brett Larsen3
1SCIEX, Vaughan, Ontario, L4K 4V8, Canada; Lunenfeld-Tanenbaum Research Institute, Mount Sinai Hospital, Sinai Health, Toronto, Ontario, M5G 1X5, Canada; NHC Key Laboratory of Human Stem Cell and Reproductive Engineering, School of Basic Medical Sciences, Central South University, Changsha, Hunan, 410075, China; Clinical Research Center for Reproduction and Genetics in Hunan Province, Reproductive and Genetic Hospital of CITIC-XIANGYA, Changsha, Hunan, 410000, China.
Background:
Mass spectrometry (MS)-based proteomics is a powerful tool for identifying and quantifying proteins. However, chimeric spectra caused by the fragmentation of multiple precursors within the same isolation window impair the accuracy of peptide identification and isobaric mass tag-based quantification. While there have been advances in computational deconvolution of chimeric spectra and methods to further separate the peptides by ion mobility or through MSn, the use of narrower isolation windows to decrease the fraction of chimeric species remains to be fully explored.
Results:
We present results obtained on a SCIEX TripleTOF instrument where the quadrupole was optimized and tuned for precursor isolation at 0.1 Da (FWHH). Using a three-proteome model (trypsin digest of protein lysates from yeast, human and E. coli) and 8-plex iTRAQ labeling to document the interference effect, we investigated the impact of co-fragmentation on spectral purity, identification accuracy and quantification accuracy. The narrow quadrupole isolation window significantly improved the spectral purity and reduced the interference of non-target precursors on quantification accuracy. The high-resolution isolation strategy also reduced the number of false identifications caused by chimeric spectra. While these improvements came at the cost of sensitivity loss, combining high-resolution isolation with other advanced techniques, including ion mobility, may result in improved accuracy in identification and quantification.
Significance:
Compared to standard-resolution quadrupole isolation (0.7 Da), high-resolution quadrupole isolation (0.1 Da) significantly improved the spectral purity and quantification accuracy while reducing the number of potential false identifications caused by chimeric spectra, thus showing excellent potential for further development to analyze clinical proteomics samples, for which high accuracy is essential.
More Related Videos
10:37Deep Proteome Profiling by Isobaric Labeling, Extensive Liquid Chromatography, Mass Spectrometry, and Software-assisted Quantification
Published on: November 15, 2017
11:49Quantitative Proteomics Workflow using Multiple Reaction Monitoring Based Detection of Proteins from Human Brain Tissue
Published on: August 28, 2021
Related Concept Videos
Mass Analyzers: Common Types
High-Resolution Mass Spectrometry (HRMS)
Mass Analyzers: Overview
Peptide Identification Using Tandem Mass Spectrometry
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...
¹³C NMR: Distortionless Enhancement by Polarization Transfer (DEPT)
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...