Related Experiment Video
Updated: Jun 13, 2025

06:33
Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
11.3K
Measurement of computed tomography modulation transfer function with a novel polymethyl methacrylate phantom
1X-ray Safety Australia, Ashgrove West, PO Box 574, Ashgrove West, QLD, 4060, Australia. jack@xraysafety.com.au.
Physical and Engineering Sciences in Medicine
|September 9, 2024
Summary
A new phantom made from polymethyl methacrylate (PMMA) can measure the modulation transfer function (MTF) for computed tomography (CT) scanners. This tool offers a cost-effective method for routine CT scanner quality control testing.
Area of Science:
- Medical Imaging Physics
- Radiological Technology
- Quality Assurance in Medical Imaging
Background:
- Accurate measurement of the modulation transfer function (MTF) is crucial for assessing computed tomography (CT) scanner performance.
- Existing methods for MTF measurement can be complex or costly, limiting routine quality control (QC).
Purpose of the Study:
- To investigate a novel, cost-effective phantom for measuring the 10% and 50% MTF values of CT scanners.
- To evaluate the accuracy, reproducibility, and applicability of the novel phantom for routine CT QC.
Main Methods:
- A novel phantom was constructed using polymethyl methacrylate (PMMA) with drilled rows of holes of varying sizes and frequencies.
- MTF was determined by analyzing Hounsfield unit ranges within the holes relative to air and PMMA.
- Results were compared against a conventional thin-wire method using identical acquisition and reconstruction parameters.
Main Results:
- The drilled hole phantom demonstrated reasonable accuracy for 50% MTF measurements.
- The phantom underestimated the 10% MTF by an average of 8.2% compared to the conventional method.
- Measurements were reproducible across repeated acquisitions and different users, and accurately reflected changes with different reconstruction kernels.
Conclusions:
- The novel drilled hole phantom is a viable, inexpensive, and user-friendly tool for routine CT scanner QC testing.
- While accurate for 50% MTF, further refinement may be needed for precise 10% MTF measurements.
- The phantom's reproducibility and sensitivity to reconstruction kernels support its application in clinical settings.

