Impact of commissural alignment on the hemodynamic performance of supra-annular self-expandable transcatheter aortic

Ignacio J Amat-Santos1,2, Javier Gómez-Herrero1, Pablo Pinon3

  • 1Cardiology Department, University Clinic Hospital, Valladolid, Spain.

Insights

Commissural alignment (CA) strategies reduced commissural misalignment in self-expandable transcatheter aortic valve replacement (TAVR), leading to lower transvalvular gradients and regurgitation progression at one year.

Area of Science:

  • Cardiovascular Medicine
  • Interventional Cardiology
  • Biomedical Engineering

Background:

  • The hemodynamic impact of commissural alignment (CA) in self-expandable transcatheter aortic valve replacement (TAVR) remains under-investigated.
  • Optimizing TAVR prosthesis deployment is crucial for long-term hemodynamic performance.

Purpose of the Study:

  • To evaluate the hemodynamic consequences of commissural alignment (CA) strategies during self-expandable TAVR.
  • To assess the impact of CA on transvalvular gradients and regurgitation post-TAVR.

Main Methods:

  • A multicentric ambispective study compared patients receiving Evolut Pro/Pro+ (EP) and Acurate neo2 (AN2) TAVR with or without CA strategies.
  • Commissural misalignment (CMA) was quantified using CT/angiography, and 1-year hemodynamic outcomes (gradients, regurgitation) were assessed via echocardiography.
  • Matched comparisons were performed based on annular dimensions, prosthesis characteristics, and baseline patient data.

Main Results:

  • CA strategies successfully reduced the incidence of moderate/severe commissural misalignment (CMA) compared to no CA (73.5% vs. 43.6%).
  • Patients with moderate/severe CMA exhibited higher peak and mean transvalvular gradients at 1-year post-TAVR.
  • A higher rate of central aortic regurgitation was observed in patients with moderate/severe CMA, though structural deterioration criteria were not met.

Conclusions:

  • Commissural alignment strategies effectively minimize commissural misalignment in self-expandable TAVR devices (AN2 and EP).
  • Reduced CMA is associated with favorable 1-year hemodynamic outcomes, including lower transvalvular gradients and less intra-prosthetic regurgitation progression.
  • CA strategies appear beneficial for optimizing TAVR hemodynamic performance without immediate signs of structural valve deterioration.
Abstract