Assessing the diagnostic accuracy of artificial intelligence in post-endovascular aneurysm repair endoleak detection
Ewa Nowak1, Marcin Białecki1,2, Agnieszka Białecka3
1Department of Radiology and Diagnostic Imaging, Collegium Medicum, Nicolaus Copernicus University in Toruń, Bydgoszcz, Poland.
Purpose:
The aim of this study was to evaluate the diagnostic accuracy of an artificial intelligence (AI) tool in detecting endoleaks in patients undergoing endovascular aneurysm repair (EVAR) using dual-energy computed tomography angiography (CTA).
Material And Methods:
The study involved 95 patients who underwent EVAR and subsequent CTA follow-up. Dualenergy scans were performed, and images were reconstructed as linearly blended (LB) and 40 keV virtual monoenergetic (VMI) images. The AI tool PRAEVAorta®2 was used to assess arterial phase images for endoleaks. Two experienced readers independently evaluated the same images, and their consensus served as the reference standard. Key metrics, including accuracy, precision, recall, F1 score, and area under the receiver operating characteristic (ROC) curve (AUC), were calculated.
Results:
The final analysis included 94 patients. The AI tool demonstrated an accuracy of 78.7%, precision of 67.6%, recall of 10 71.9%, F1 score of 69.7%, and an AUC of 0.77 using LB images. However, the tool failed to process 40 keV VMI images correctly, limiting further analysis of these datasets.
Conclusions:
The AI tool showed moderate diagnostic accuracy in detecting endoleaks using LB images but failed to achieve the reliability needed for clinical use due to the significant number of misdiagnoses.
More Related Videos
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
07:12Author Spotlight: Using Point-of-Care Ultrasound for Comprehensive Evaluation of the Abdominal Aorta
Published on: September 8, 2023
Related Concept Videos
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for diagnosing...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Peripheral Arterial Disease II: Clinical Manifestations and Diagnostic Evaluation
Venous Thrombosis II: Clinical Manifestations and Diagnostic Studies