Spectral transmission of commercially available high-energy visible light-filtering spectacle lenses
Nishanee Rampersad, Anthony Carlson1
1Department of Optometry, University of Johannesburg, Johannesburg, South Africa.
Significance:
High-energy visible (HEV) light-filtering spectacle lenses are being widely promoted despite limited evidence to support their use. The spectral transmission properties of commercially available lenses varied, particularly in the HEV light region, and they do not perform in the same way for transmission of ultraviolet (UV)-A and HEV light.
Purpose:
Although HEV light is important for several visual and nonvisual functions, there are concerns over the potential adverse effects of increased HEV light exposure. High-energy visible light-filtering spectacle lenses are being widely marketed to promote ocular health and improve sleep by reducing exposure to HEV light. This study reports on the spectral transmission characteristics of commercially available HEV light-filtering spectacle lenses based on the recommendations of the Spectral Bands Task Force technical report.
Methods:
The spectral transmission of light through nine afocal plastic lenses, including eight commercially available HEV light-filtering lenses and one clear uncoated control lens, was evaluated using a Cary 5000 UV-Vis-NIR spectrophotometer (Agilent Technologies, Johannesburg, South Africa) for wavelengths 250 to 780 nm. The percentage transmission values are reported for UV radiation and visible light, with emphasis for HEV light (380 to 500 nm) and the three subbands therein.
Results:
All lenses blocked UV-C and UV-B radiation (250 to 315 nm). For UV-A radiation (315 to 380 nm), six lenses showed optimal 100% absorption, whereas three lenses allowed ≤12%. The transmission values for the HEV light-filtering lenses ranged from 55 to 90% and 75 to 95% for HEV light of wavelengths 400 to 455 nm and 455 to 500 nm, respectively. In contrast, the control lens showed 92 (400 to 455 nm) and 93% (455 to 500 nm) transmission.
Conclusions:
The HEV light-filtering spectacle lenses varied in their transmission properties and do not necessarily perform in the same way for transmission of UV-A and HEV light. Optometric personnel should consider these transmission properties, particularly for HEV light, when making recommendations to patients about HEV light-filtering spectacle lenses.
More Related Videos
06:55Scanning Light Scattering Profiler SLPS Based Methodology to Quantitatively Evaluate Forward and Backward Light Scattering from Intraocular Lenses
Published on: June 6, 2017
08:33Determination of Photoreceptor Cell Spectral Sensitivity in an Insect Model from In Vivo Intracellular Recordings
Published on: February 26, 2016
Related Concept Videos
UV–Vis Spectrometers
Photoreceptors and Visual Pathways
Ultraviolet and Visible (UV–Vis) Spectroscopy: Overview
UV–Vis Spectroscopy: Molecular Electronic Transitions
UV–Vis Spectroscopy of Conjugated Systems
One of the factors influencing λmax is the extent...
Energy Transfer in Chemical Reactions