Correction Factor Evaluation and Between-System Comparison of Behavioral Threshold Predictions From Auditory
Marlene Bagatto1, Rana El-Naji1, David Purcell1
1National Centre for Audiology, Western University, London, Ontario, Canada.
American Journal of Audiology
|September 11, 2024
Summary
Auditory brainstem response (ABR) correction factors accurately predict infant hearing thresholds. Threshold estimates are comparable between different ABR systems when parameters are similar, ensuring reliable hearing assessments.
Area of Science:
- Audiology
- Pediatric audiology
- Hearing science
Background:
- Accurate hearing threshold estimation in infants is crucial for early intervention.
- Auditory Brainstem Response (ABR) is a common objective measure used in infants.
- Corrections are applied to ABR thresholds to estimate behavioral thresholds.
Purpose of the Study:
- To validate the accuracy of ABR correction factors used in the Ontario Infant Hearing Program.
- To compare behavioral threshold estimates between different ABR systems.
- To inform equipment transition and protocols in Ontario, Canada.
Main Methods:
- Study 1: Retrospective file review comparing ABR threshold estimates to behavioral thresholds in 84 infants with hearing loss.
- Study 2: Examination of the precision of two ABR systems in estimating thresholds in 37 adult and 105 infant ears.
- Statistical analysis to determine the accuracy of ABR correction factors and differences between systems.
Main Results:
- Corrected ABR thresholds accurately predicted behavioral thresholds in infants, with an average difference of 1.77 dB.
- Accounting for ear canal acoustics reduced the average difference to 0.6 dB.
- The average difference between two ABR systems was 2.40 dB, indicating comparable threshold estimates.
Conclusions:
- The ABR correction factors used in Ontario's Infant Hearing Program reliably predict behavioral thresholds in infants.
- Threshold estimates are comparable across different ABR systems if calibration and collection parameters are similar.
- No further adjustment to correction factors is needed when using comparable ABR systems.