Jove
Visualize
Contact Us
JoVE
x logofacebook logolinkedin logoyoutube logo
ABOUT JoVE
OverviewLeadershipBlogJoVE Help Center
AUTHORS
Publishing ProcessEditorial BoardScope & PoliciesPeer ReviewFAQSubmit
LIBRARIANS
TestimonialsSubscriptionsAccessResourcesLibrary Advisory BoardFAQ
RESEARCH
JoVE JournalMethods CollectionsJoVE Encyclopedia of ExperimentsArchive
EDUCATION
JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab ManualFaculty Resource CenterFaculty Site
Terms & Conditions of Use
Privacy Policy
Policies

Related Concept Videos

Magnetic Resonance Imaging01:24

Magnetic Resonance Imaging

5.0K
Magnetic resonance imaging (MRI) is a noninvasive medical imaging technique based on a phenomenon of nuclear physics discovered in the 1930s, in which matter exposed to magnetic fields and radio waves was found to emit radio signals. In 1970, a physician and researcher named Raymond Damadian noticed that malignant (cancerous) tissue gave off different signals than normal body tissue. He applied for a patent for the first MRI scanning device in clinical use by the early 1980s. The early MRI...
5.0K
Fast Decoupled and DC Powerflow01:24

Fast Decoupled and DC Powerflow

178
The fast decoupled power flow method addresses contingencies in power system operations, such as generator outages or transmission line failures. This method provides quick power flow solutions, essential for real-time system adjustments. Fast decoupled power flow algorithms simplify the Jacobian matrix by neglecting certain elements, leading to two sets of decoupled equations:
178
Double Resonance Techniques: Overview01:12

Double Resonance Techniques: Overview

191
Double resonance techniques in Nuclear Magnetic Resonance (NMR) spectroscopy involve the simultaneous application of two different frequencies or radiofrequency pulses to manipulate and observe two distinct nuclear spins. One important application of double resonance is spin decoupling, which selectively suppresses coupling with one type of nucleus while observing the NMR signal from another nucleus, simplifying the spectrum and enhancing resolution.
Spin decoupling is usually achieved by...
191

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Effect of Medicaid coverage of tobacco-dependence treatments on smoking cessation.

International journal of environmental research and public health·2010
Same author

Cytokine and autoantibody patterns in acute liver failure.

Journal of immunotoxicology·2009
Same author

A novel scoring system for prognostic prediction in d-galactosamine/lipopolysaccharide-induced fulminant hepatic failure BALB/c mice.

BMC gastroenterology·2009
Same author

Mammalian target of rapamycin signaling pathway contributes to glioma progression and patients' prognosis.

The Journal of surgical research·2009
Same author

Estrogen receptor neurobiology and its potential for translation into broad spectrum therapeutics for CNS disorders.

Current molecular pharmacology·2009
Same author

Transcriptional and post-translational regulation of adiponectin.

The Biochemical journal·2009

Related Experiment Video

Updated: Jun 13, 2025

Quantitative Magnetic Resonance Imaging of Skeletal Muscle Disease
09:30

Quantitative Magnetic Resonance Imaging of Skeletal Muscle Disease

Published on: December 18, 2016

19.5K

A hybrid, nonlinear programming approach for optimizing passive shimming in MRI.

Jie Zhao1, Minhua Zhu1, Ling Xia2

  • 1School of Medical Imaging, Hangzhou Medical College, Hangzhou, China.

Medical Physics
|September 13, 2024
PubMed
Summary

This study introduces a hybrid algorithm for magnetic resonance imaging (MRI) passive shimming, significantly improving magnetic field uniformity and reducing iron usage. The new method optimizes shimming for better image quality and efficiency.

Keywords:
hybrid optimizationmagnetic resonance imagingparticle swarm optimizationpassive shimmingsequential quadratic programming

More Related Videos

Cardiac Magnetic Resonance Imaging at 7 Tesla
09:14

Cardiac Magnetic Resonance Imaging at 7 Tesla

Published on: January 6, 2019

11.4K
Registered Bioimaging of Nanomaterials for Diagnostic and Therapeutic Monitoring
17:16

Registered Bioimaging of Nanomaterials for Diagnostic and Therapeutic Monitoring

Published on: December 9, 2010

10.3K

Related Experiment Videos

Last Updated: Jun 13, 2025

Quantitative Magnetic Resonance Imaging of Skeletal Muscle Disease
09:30

Quantitative Magnetic Resonance Imaging of Skeletal Muscle Disease

Published on: December 18, 2016

19.5K
Cardiac Magnetic Resonance Imaging at 7 Tesla
09:14

Cardiac Magnetic Resonance Imaging at 7 Tesla

Published on: January 6, 2019

11.4K
Registered Bioimaging of Nanomaterials for Diagnostic and Therapeutic Monitoring
17:16

Registered Bioimaging of Nanomaterials for Diagnostic and Therapeutic Monitoring

Published on: December 9, 2010

10.3K

Area of Science:

  • Medical Imaging
  • Physics
  • Engineering

Background:

  • Magnetic Resonance Imaging (MRI) requires a highly uniform main magnetic field (B0) for detailed anatomical imaging.
  • Passive shimming (PS) uses iron pieces to improve B0 uniformity, but traditional linear programming (LP) methods struggle to balance field quality with iron quantity.
  • Existing PS optimization faces challenges in achieving optimal field quality while minimizing iron usage.

Purpose of the Study:

  • To enhance passive shimming efficacy by balancing field quality, iron usage, and harmonics for a smoother magnetic field profile.
  • Develop an improved passive shimming technique for MRI magnets.
  • Achieve optimal magnetic field uniformity with reduced iron material.

Main Methods:

  • Introduced a hybrid algorithm combining particle swarm optimization (PSO) with sequential quadratic programming (SQP) for enhanced shimming performance.
  • Employed a regularization method to effectively reduce the weight of iron pieces used in shimming.
  • Utilized a novel computational approach for passive shimming optimization.

Main Results:

  • Achieved a significant improvement in magnetic field uniformity, reducing field inhomogeneity from 462 ppm to 3.6 ppm within a 40 cm diameter spherical volume of a 7T MRI magnet.
  • Demonstrated a 96.7% enhancement in magnetic field uniformity compared to traditional methods.
  • Reduced the required iron weight by 81.8%, using only 1.2 kg of material.

Conclusions:

  • The proposed hybrid PSO-SQP algorithm with regularization is effective for passive shimming in MRI.
  • This method offers a promising approach for optimizing magnetic field uniformity and material efficiency in MRI systems.
  • The findings suggest potential for improved image quality and reduced construction costs in MRI technology.