Case:
A 34-year-old man with a distal tibia bone defect was treated with an intramedullary bone transport nail (Precice Bone Transport System, NuVasive). During planned removal after successful treatment, 7 separate subcomponents of the nail became disconnected and had to be separately removed using specialized instrumentation. This occurred despite adherence to the manufacturer's recommended technique for nail removal and in the absence of clinical or radiographic evidence of implant failure.
Conclusion:
When planning for implant removal, surgeons should be aware of potential intraoperative disconnection of subcomponents of this magnetic bone transport nail and ensure that equipment for retrieval (e.g., very long endoscopy forceps) is available.
More Related Videos
07:35Creating Rigidly Stabilized Fractures for Assessing Intramembranous Ossification, Distraction Osteogenesis, or Healing of Critical Sized Defects
Published on: April 11, 2012
06:41A Minimally Invasive Model to Analyze Endochondral Fracture Healing in Mice Under Standardized Biomechanical Conditions
Published on: March 22, 2018
Related Concept Videos
Fractures: Bone Repair
Minor fractures with no bone displacement are treated by immobilizing the fractured bone using a cast or splint. However, in the case of fractures with displaced bones, the broken bones are repositioned before immobilization to ensure successful healing without deformation and loss of function. The realignment of fractured bone ends is performed through a process called reduction. If the...
Bone Remodeling