Development of Robust PEBAX-Based Angiographic Catheter: Design and In Vitro Study

Hafsa Inam1, Murtaza Najabat Ali1, Ibraheem Raza Jameel2

  • 1Biomedical Engineering and Sciences Department, School of Mechanical and Manufacturing Engineering (SMME), National University of Sciences and Technology (NUST), Islamabad 44000, Pakistan.

PubMed

Insights

A new PEBAX®-based angiographic catheter demonstrated superior performance, significantly reducing advancement force and vascular spasm during diagnostic procedures. This innovative catheter offers enhanced flow for radiopaque material delivery in coronary arteries.

Area of Science:

  • Cardiovascular diagnostics
  • Medical device engineering
  • Materials science in medicine

Background:

  • Coronary disease prevalence necessitates improved diagnostic tools.
  • Angiography is the gold standard, using amide and urethane catheters.
  • PEBAX®'s potential benefits in angiography are under-explored.

Purpose of the Study:

  • Develop a PEBAX®-based angiographic catheter.
  • Evaluate its performance against commercial nylon and polyurethane catheters.

Main Methods:

  • Developed a PEBAX®-based catheter.
  • Assessed pushability (advancement force) and arterial force endurance.
  • Conducted bench tests per regulatory standards.

Main Results:

  • PEBAX® catheter showed lowest bond-flexural rigidity (180.4 g).
  • PEBAX® catheter exhibited 50% less advancement force (510.4 g) than nylon/polyurethane.
  • Demonstrated superior performance in a mock arterial system.

Conclusions:

  • PEBAX® catheters outperformed nylon and polyurethane types.
  • Minimal advancement force suggests enhanced radiopaque material delivery.
  • Potential to significantly minimize vascular spasms during diagnostic angiography.