Related Experiment Video
Updated: Jun 13, 2025

Manufacturing Abdominal Aorta Hydrogel Tissue-Mimicking Phantoms for Ultrasound Elastography Validation
Published on: September 19, 2018
Development of a mechanical characterisation device for intracranial aneurysms: Calibration on polymeric phantom
G Plet1, J Raviol1, H Magoariec1
1Laboratoire de Tribologie et Dynamique des Systèmes, CNRS UMR 5513, Université de Lyon, Ecole Centrale de Lyon, France.
Abstract:
Intracranial aneurysm is a major health issue related to biomechanical arterial wall degradation. Currently, no method allows predicting rupture risk based on in vivo quantitative mechanical data. This work is part of a large-scale project aimed at providing clinicians with a non-invasive patient-specific decision support tool, based on the in vivo mechanical characterisation of the aneurysm wall. Thus, the primary objective of the project was to develop a deformation device prototype (DDP) of the artery wall and to calibrate it on polymeric phantom arteries. The deformations induced on the phantom arteries were quantified experimentally using a Digital Image Correlation (DIC) system. The results indicated that the DIC system was able to measure the small displacements generated by the DDP. We also observed that the flow mimicking the blood flow did not significantly disturb the measurements of the artery wall displacement caused by the DDP. Finally, a limit displacement value generated by the DDP was evaluated. This value corresponds to the lowest displacement value detectable by the clinical imaging system that will be tested on animals in the future (Spectral Photon Counting CT).

