How differ eCAP types in cochlear implants users with and without inner ear malformations: amplitude growth function,
Betul Cicek Cinar1, Merve Özses2
1Department of Audiology, Hacettepe University, Ankara, Turkey. betulcicek01@gmail.com.
Inner ear malformations (IEMs) impact cochlear implant (CI) outcomes. While the presence rate of electrically evoked compound action potentials (eCAPs) varied, IEM and normal groups showed similar eCAP amplitudes, suggesting comparable neural responses despite structural differences.
Area of Science:
- Otolaryngology
- Neuroscience
- Biomedical Engineering
Background:
- Inner ear malformations (IEMs) can influence cochlear implant (CI) outcomes.
- Electrically evoked compound action potentials (eCAPs) are key objective measures in CI users.
- eCAP presence rates differ between CI users with and without IEMs.
Purpose of the Study:
- To evaluate eCAP results in CI users with and without IEMs.
- To compare different eCAP recording methods: amplitude growth function (AGF), spreads of excitation (SOE), and refractory recovery (RecF).
- To assess the impact of IEMs on eCAP presence, thresholds, and amplitudes.
Main Methods:
- Studied 42 CI users (20 with IEMs, 22 normal), aged over five years with at least one year of CI experience.
- Conducted three eCAP measurements (AGF, SOE, RecF) at multiple intracochlear electrodes.
- Compared eCAP presence rates, threshold levels, and amplitudes between the IEM and normal groups.
Main Results:
- A significant difference in eCAP threshold detection was found only at the basal electrode.
- No significant differences in AGF-eCAP and RecF-eCAP amplitudes were observed across apical, middle, and basal regions between groups.
- While SOE-eCAP presence was lower in the IEM group, eCAP amplitudes remained similar for all maskers.
Conclusions:
- Observable eCAP rates differ between CI users with and without IEMs.
- Despite differing presence rates, CI users with IEMs demonstrate eCAP amplitudes comparable to those with normal cochlear structures.
- IEMs may not significantly alter the neural response amplitude measured by eCAPs.