Related Experiment Videos
Serial determination of lung volume in small animals by nitrogen washout
Journal of Applied Physiology (Bethesda, Md. : 1985)
|July 1, 1985
Abstract:
This report describes the design of an apparatus and the procedures used to serially measure the total lung capacity and the functional residual capacity of small animals utilizing the N2-washout technique. The calibration data indicate that the technique is accurate to within 1 ml and has a variance of less than 5%. The in vivo lung volume measurements of rats were validated by comparing them with values obtained with a water-displacement technique; the means were within 0.3 ml. Examples of the precision and changes in lung volume of animals during studies are included to demonstrate the reliability and usefulness of the technique.