Evaluating the accuracy of planar gated blood pool processing software using simulated patient studies.
H Pieters1, J A van Staden2, H du Raan2
1Nuclear Medicine Unit, Klerksdorp/Tshepong Hospital Complex, Klerksdorp, 2571, South Africa.
This study validates commercial software for assessing left ventricular ejection fraction (LVEF) using simulated cardiac imaging. While generally accurate and reproducible, some underestimation of LVEF occurred at lower values.
Area of Science:
- Nuclear Medicine
- Cardiovascular Imaging
- Medical Physics
Background:
- Planar gated blood pool (GBP-P) imaging is crucial for non-invasive assessment of left ventricular ejection fraction (LVEF).
- Accurate LVEF monitoring is vital for patients undergoing cardiotoxic chemotherapy.
- Concerns exist regarding the accuracy of commercial software for LVEF determination despite improved reproducibility.
Purpose of the Study:
- To create a database of GBP-P studies with known LVEF values using Monte Carlo simulations.
- To evaluate the accuracy of LVEF measurements from four commercial software programs.
- To assess the reproducibility of LVEF values derived from simulated GBP-P studies.
Main Methods:
- Utilized anthropomorphic 4D-XCAT models to generate 64 realistic GBP-P studies via Monte Carlo simulations.
- Processed simulated studies using four commercial software packages: Alfanuclear, Siemens, GE Xeleris, and Mediso Tera-Tomo.
- Assessed accuracy, intra-observer, and inter-observer reproducibility of LVEF measurements.
Main Results:
- Demonstrated a strong correlation between software-calculated LVEF and true LVEF values from simulations.
- Observed a slight underestimation of LVEF by all software programs at lower LVEF values.
- Reported excellent intra- and inter-observer reliability for LVEF measurements, indicating reduced subjectivity.
Conclusions:
- Commercial software programs accurately and reliably determine LVEF from simulated GBP-P studies.
- The observed underestimation bias at low LVEF necessitates careful interpretation to avoid excluding chemotherapy candidates.
- Future research should focus on mitigating underestimation biases and validating findings across diverse patient populations.