Normative data for macular perimetry using the MP-3 microperimeter in healthy individuals
Taurino Dos Santos Rodrigues Neto1, Epitácio Dias da Silva Neto1, Alex Haruo Higashi1
1Division of Ophthalmology, Laboratory for Investigation in Ophthalmology (LIM-33), Faculdade de Medicina, Universidade de São Paulo, São Paulo, SP, Brazil.
Arquivos Brasileiros De Oftalmologia
|September 19, 2024
Summary
This study establishes normal microperimetry values using the MP-3 device in healthy individuals. These findings provide crucial age-matched data for assessing retinal sensitivity and fixation stability.
Area of Science:
- Ophthalmology
- Visual Neuroscience
Background:
- Microperimetry is vital for assessing visual function in retinal diseases.
- Establishing normal microperimetry values with the MP-3 is essential for accurate impairment assessment.
- Baseline data for macular sensitivity, age, and sex correlations are needed.
Purpose of the Study:
- Determine normal light sensitivity thresholds using the MP-3 microperimeter in healthy individuals.
- Assess fixation stability during microperimetry with the MP-3.
- Correlate microperimetry values with age and sex.
Main Methods:
- Full-threshold microperimetry with a 4-2 staircase strategy was performed on 37 healthy volunteers (age 28-68).
- Standard Goldmann III stimulus size and 68 test points (similar to Humphrey Field Analyzer 10-2) were used.
- Fixation stability was recorded simultaneously, and age-related sensitivity decline was analyzed via linear regression.
Main Results:
- Global mean sensitivity was 29.01 ± 1.44 dB.
- Mean central sensitivity at 2° was 28.5 ± 1.77 dB (OD) and 28.75 ± 1.98 dB (OS).
- Median fixation stability within 2° and 4° was 80% and 96%, respectively. A decline in global sensitivity with age was observed.
Conclusions:
- MP-3 microperimetry provides accurate, topography-specific retinal sensitivity testing.
- This study offers a normal, age-matched database for MP-3 microperimetry.
- The data aids in evaluating retinal function and diagnosing visual impairments.
Related Concept Videos
Measurements of Strain
Strain quantifies the deformation of a material under force, typically measured as normal strain, which represents the change in length when compared with the original length. Electrical strain gauges are used for enhanced accuracy. These devices consist of a conductive wire mounted on a paper backing that adheres to the material's surface. These gauges operate on the piezoresistive effect, where the wire's electrical resistance changes in response to mechanical deformation. The strain gauge...
Fineness Modulus
The fineness modulus (FM) of aggregate is a numerical index that measures the coarseness or fineness of the particles. It is calculated by adding the cumulative percentages of aggregate retained on each of a specified series of sieves and dividing the sum by 100.
Consider performing sieve analysis on sand through a set of ASTM sieves. The weight of aggregate retained in each sieve and pan placed at the bottom is recorded, as given in Column B of Table 1.
To determine the fineness modulus of...
Consider performing sieve analysis on sand through a set of ASTM sieves. The weight of aggregate retained in each sieve and pan placed at the bottom is recorded, as given in Column B of Table 1.
To determine the fineness modulus of...