Radiation exposure during simulated equine head and limb fan beam standing computed tomography appears safe for
Kaylynn E Veitch1, Timothy P Szczykutowicz1,2, Sabrina H Brounts1
11Department of Surgical Sciences, School of Veterinary Medicine, University of Wisconsin-Madison, Madison, WI.
Objective:
To evaluate the radiation dose to personnel locations during simulated head and limb scans with a novel equine standing CT (sCT) system.
Methods:
Measurements were made with the use of a helical fan beam sCT system (Equina; Asto CT Inc). Scatter radiation was measured in different positions in the sCT room to mimic the location of the control operator, horse handler, and lead rope handler during simulated equine head and limb imaging. Operator/handler dose was quantified at each location using entrance air kerma measured with a spherical ionization chamber and electrometer.
Results:
Radiation dose to the control operator, horse handler, and lead rope handler locations wearing a lead apron during simulated head imaging was 13.3, 3.5, and 6.8 µGy, respectively. Radiation dose to the control operator location wearing a lead apron was 1.3 µGy, and dose to the lead rope handler location wearing a lead apron was 0.2 and 5.4 µGy during simulated pelvic limb and thoracic limb imaging, respectively.
Conclusions:
With the more widespread clinical use of equine sCT units in clinical practice, there is concern for increased risk of radiation exposure to personnel who stay in the sCT room during scanning. The control operator location had the highest dose during simulated head imaging, and the lead rope handler location in thoracic limb sCT had the highest dose during simulated limb imaging. Limiting the number of personnel in the sCT room, rotating personnel between handler positions, increasing operator distance from the scanner, and using lead shields and eyeglasses are recommended.
Clinical Relevance:
Our findings suggest that scanning large numbers of horses per year with the Asto CT Equina would not lead to occupational radiation exposure above the recommended safe threshold for handlers using lead shields and eyeglasses.
More Related Videos
09:07Multi-Tracer Studies of Brain Oxygen and Glucose Metabolism Using a Time-of-Flight Positron Emission Tomography-Computed Tomography Scanner
Published on: June 7, 2024
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...