Proximal Cadaveric Femur Preparation for Fracture Strength Testing and Quantitative CT-based Finite Element Analysis
Published on: March 11, 2017
Bone support correlation of X-Ray and CT for a new PE-glenoid
Matthias Bülhoff1, Nikolai Sonntag2, Raphael Trefzer2
1Department of Orthopaedics, University Hospital Heidelberg, Schlierbacher Landstrasse 200a, Heidelberg, 69118, Germany. matthias.buelhoff@med.uni-heidelberg.de.
Introduction:
The radiographic evaluation of novel cementless anatomic polyethylene (PE) glenoid components featuring a titanium-coated back is still unclear. This study explores potential radiolucent lines (RLL) between the radiopaque titanium layer and sclerotic convex reamed bone in an intermodal comparison analysis with computed tomography (CT) scans.
Materials And Methods:
Eight RM pressfit vitamys glenoids (Mathys®) were implanted into cadaveric scapulae. In the CT scans, glenoids were quantified by evaluating ideal complete bony support (NO GAP) and gap between bone and titanium coating (GAP). X-rays were in perfect 0-degree projection and tilted in ± 10° and ± 20° mediolateral (ml) and craniocaudal (cc) directions. Radiographs evaluated were graded as NO RLL, RLL (gap > 1 mm) or DL (double line, gap < 1 mm) in an intermodal comparison of CT and X-ray findings.
Results:
The inter-rater (Cohen's = 0.643) and intra-rater reliability (Cohen's = 0.714) were good. The overall evaluation showed a significant agreement between (NO) RLL on X-ray and (NO) GAP on CT (p < 0.001). The - 10-degree ml projection showed good agreement between CT and X-ray (Cohen's = 0.628). Adequate agreement was shown at 0 degrees (Cohen's = 0.386), + 10 degrees ml (Cohen's = 0.338), and + 20 degrees cc (Cohen's = 0.327). Compared to the scenario DL = NO RLL, the true a.p. view showed better sensitivity when the DL is classified as RLL. Conversely, the true a.p. view demonstrated both better specificity and significant agreement between the X-ray and CT findings in scenario when DL = No RLL.
Conclusion:
Standard true a. p. projections are reliable in ruling out gaps when no RLL or DL is visible and the detection of RLL shows high intermodal agreement. Varying agreement across tilting angles emphasizes the importance of a comprehensive approach in evaluating bone support and CT is indispensable for a scientifically reliable assessment.
Level Of Evidence:
Level III Treatment Study.
Related Concept Videos
Radiological Investigation I: X-ray and CT
Gross Anatomy of Bone
The diaphysis is the tubular shaft that runs between the proximal and distal ends of the bone. The walls of the diaphysis are composed of dense and hard compact bone made of numerous osteons — the functional unit of the compact bone. The hollow region in the diaphysis is called the medullary cavity, which harbors the bone marrow. In infants and children, this marrow cavity is filled with red marrow, whereas in...
Bones of the Upper Limb: Humerus
Bone Structure
Classification of Bones
Long and Short Bones
The appendicular skeleton, particularly the upper and lower limbs, is primarily made of long and short bones. The...
Bones of the Upper Limb: Radius
The radius has a nail-shaped head, and a...