Novel Quantitative Liver Steatosis Assessment Method With Ultrasound Harmonic Imaging
Ping Gong1, Jingke Zhang1, Chengwu Huang1
1Department of Radiology, Mayo Clinic, Rochester, Minnesota, USA.
Objectives:
Metabolic dysfunction-associated steatotic liver disease (MASLD) is the most prevalent liver disorder in Western countries, with approximately 20%-30% of the MASLD patients progressing to severe stages. There is an urgent need for noninvasive, cost-effective, widely accessible, and precise biomarkers to evaluate liver steatosis. This study aims to assess and compare the diagnostic performance of a novel reference frequency method-based ultrasound attenuation coefficient estimation (ACE) in both fundamental (RFM-ACE-FI) and harmonic (RFM-ACE-HI) imaging for detecting and grading liver steatosis.
Methods:
An Institutional Review Board-approved prospective study was carried out between December 2018 and October 2022. A total number of 130 subjects were enrolled in the study. The correlation between RFM-ACE-HI values and magnetic resonance imaging proton density fat fraction (MRI-PDFF), as well as between RFM-ACE-FI values and MRI-PDFF were calculated. The diagnostic performance of RFM-ACE-FI and RFM-ACE-HI was evaluated using receiver operating characteristic (ROC) curve analysis, as compared to MRI-PDFF. The reproducibility of RFM-ACE-HI was assessed by interobserver agreement between two sonographers.
Results:
A strong correlation was observed between RFM-ACE-HI and MRI-PDFF, with R = 0.88 (95% confidence interval [CI]: 0.83-0.92; P < .001), while the correlation between RFM-ACE-FI and MRI-PDFF was R = 0.65 (95% CI: 0.50-0.76; P < .001). The area under the ROC (AUROC) curve for RFM-ACE-HI in staging liver steatosis grades of S ≥ 1 and S ≥ 2 was 0.97 (95% CI: 0.91-0.99; P < .001) and 0.98 (95% CI: 0.93-1.00; P < .001), respectively, and 0.76 (95% CI: 0.65-0.85) and 0.80 (95% CI: 0.70-0.88) for RFM-ACE-FI, respectively. Great reproducibility was achieved for RFM-ACE-HI, with an interobserver agreement of R = 0.97 (95% CI: 0.94-0.99; P < .001).
Conclusions:
The novel RFM-ACE-HI method offered high liver steatosis diagnostic accuracy and reproducibility, which has important clinical implications for early disease intervention and treatment evaluation.
More Related Videos
05:37Author Spotlight: A Non-Invasive Tool to Assess and Differentiate Fat Patterns in Liver Using 3D Dixon MRI
Published on: October 20, 2023
07:12Author Spotlight: Analysis of Fluorescent-Stained Lipid Droplets with 3D Reconstruction for Hepatic Steatosis Assessment
Published on: June 2, 2023
Related Concept Videos
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Ultrasound I: Abdominal Ultrasonography
Abdominal ultrasonography, commonly known as abdominal ultrasound, is a vital, non-invasive medical imaging technique widely used in healthcare.
Procedure:
This diagnostic tool allows the clinician to visually inspect internal structures within the abdomen, including vital organs such as the liver, gallbladder, pancreas, kidneys, and spleen.
The abdominal ultrasound process begins with applying a special gel to the patient's skin over the abdomen. This gel enhances the...
Ultrasonography
During an ultrasonography procedure, a handheld device called...