Ultraportable Quantitative Ultrasound for Hepatic Steatosis Assessment
Baptiste Hériard-Dubreuil1, Adrien Besson2, Jonathan Mamou3
1E-Scopics, Aix-en-Provence, France; LTS5, EPFL, Lausanne, Switzerland.
Objective:
This study aimed to evaluate the performances of quantitative ultrasound (QUS) for the detection and assessment of hepatic steatosis when implemented using an ultraportable ultrasound scanner.
Methods:
Seven established QUS parameters were investigated. Ultrasound signals were acquired using a new ultraportable ultrasound device, the Hepatoscope. The feasibility of QUS using the Hepatoscope was first assessed in vitro. Clinical reliability, accuracy and staging capabilities were evaluated in 60 patients referred to a hepatology consultation for known chronic liver disease and enrolled in a prospective clinical investigation using the controlled attenuation parameter (CAP) as ground truth.
Results:
QUS parameters showed moderate (intra-class correlation coefficient [ICC] >0.50) to excellent (ICC >0.90) reliability. Two parameters, namely Lizzi-Feleppa mid-band fit and attenuation, were both reliable (ICC = 0.89 and 0.86, respectively) and correlated with the CAP (squared Pearson correlation coefficient of R2 = 0.65 and R2 = 0.6, respectively). For steatosis detection (S0 vs. ≥S1), the two parameters yielded an area under the receiving operating characteristic curve of 0.90 and 0.86, respectively (95% confidence interval: [0.81-0.99] and [0.76-0.96], respectively).
Conclusion:
QUS can be reliably and accurately implemented on ultraportable ultrasound scanners. The combination of ultraportability and quantitative assessment of liver fat is promising for large-scale screening and monitoring of hepatic steatosis.
Related Concept Videos
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Ultrasonography
During an ultrasonography procedure, a handheld device called...
Ultrasound I: Abdominal Ultrasonography
Abdominal ultrasonography, commonly known as abdominal ultrasound, is a vital, non-invasive medical imaging technique widely used in healthcare.
Procedure:
This diagnostic tool allows the clinician to visually inspect internal structures within the abdomen, including vital organs such as the liver, gallbladder, pancreas, kidneys, and spleen.
The abdominal ultrasound process begins with applying a special gel to the patient's skin over the abdomen. This gel enhances the...