Introduction:
Photon-counting detector computed tomography (PCD-CT) represents the next generation of CT technology, offering enhanced capabilities for detecting the orientation of directional leads in deep brain stimulation (DBS). This study aims to refine PCD-CT-based lead orientation determination using an automated method applicable to devices from various manufacturers, addressing current methodological limitations and improving neurosurgical precision.
Methods:
An automated method was developed to ascertain the orientation of directional DBS leads using PCD-CT data and grayscale model fitting for devices from Boston Scientific, Medtronic, and Abbott. A phantom study was conducted to evaluate the precision and accuracy of this method, comparing it with the stripe artifact method across different lead alignments relative to the CT gantry axis.
Results:
Except for the Medtronic Sensight™ lead, where detection was occasionally unfeasible if aligned normal to the z-axis of the CT gantry, a clinically very unlikely alignment, the lead orientation could be automatically determined regardless of its position. The accuracy and precision of this automated method was comparable to those of the stripe artifact method.
Conclusion:
PCD-CT enables the automatic determination of lead orientation from leading manufacturers with an accuracy comparable to the stripe artifact method, and it offers the added benefit of being independent of the clinically occurring orientation of the head and, consequently, the lead relative to the CT gantry axis.