Related Experiment Video
Updated: Jun 12, 2025

A Method for Systematic Electrochemical and Electrophysiological Evaluation of Neural Recording Electrodes
Published on: March 3, 2014
Suppression of crosstalk in multielectrode arrays with local shielding
J R Naughton1, J A Varela2, T J Connolly3
1Department of Physics, Boston College, Chestnut Hill, MA, United States.
Abstract:
Electrical crosstalk can constrain the performance of multielectrode arrays in electro- and neurophysiology, in terms of both stimulation and recording. This is especially so at high electrode density, desirable for spatiotemporal mapping of bioelectrical signals from multiple cells. Channel interference due to crosstalk is currently only partially addressed, via continuous interleaved sampling or post-data acquisition spike sorting. Here, we show that a locally-shielded electrode architecture significantly suppresses crosstalk, and enables multi-site recording at high electrode density without the need for spike sorting. Arrays of shielded electrodes, prepared by micro- and nanofabrication techniques in a vertically-oriented coaxial geometry, demonstrate at least a 400 times improvement in spatial density over the unshielded case.
More Related Videos
08:43Combined Shuttle-Box Training with Electrophysiological Cortex Recording and Stimulation as a Tool to Study Perception and Learning
Published on: October 22, 2015
07:52Multiscale Investigations of Cortical Processing by Integrating Laminar Polytrodes and Optogenetics with Micro Electrocorticography in Rodents
Published on: May 23, 2025