Assessment of image-derived input functions from small vessels for patlak parametric imaging using total-body PET/CT
Hongmei Tang1, Yang Wu1, Zhaoting Cheng1
1Department of Nuclear Medicine, Tongji Hospital, Tongji Medical College, Huazhong University of Science and Technology, Wuhan, 430000, China.
Purpose:
The image-derived input function (IDIF) from the descending aorta has demonstrated performance comparable to arterial blood sampling while avoiding its invasive nature in parametric imaging. However, in conventional PET, large vessels may not always be within the imaging field of view (FOV). This study aims to evaluate the efficacy of dynamic parametric Ki imaging using image-derived input functions (IDIFs) extracted from various arteries, facilitated by total-body PET/CT.
Method:
Twenty-three participants underwent a 60-minute total-body [18F]FDG PET scan. Data from each subject were used to reconstruct both total-body PET images and short-axis field-of-view PET images at different bed positions, each with a 25 cm axial field-of-view (AFOV). Partial volume correction (PVC) was performed using the blurred Van Cittert iterative deconvolution. IDIFs extracted from the descending aorta, carotid artery, abdominal aorta, and iliac artery were employed for Patlak analysis. The resulting Ki images were compared using quantification indicators and subjective assessment. Linear regression analysis was conducted to examine the correlation of Ki values among IDIFs in normal organ and lesion regions of interest (ROIs).
Result:
High similarities were observed in Ki images derived from the IDIFs from the descending aorta and other arteries, with a median structural similarity index measure (SSIM) above 0.98 and a median peak signal-to-noise ratio (PSNR) above 37dB. Linear regression analysis revealed strong correlations in Ki values (r² > 0.88) between the descending aorta and the three alternative vessels, with slopes of the linear fits close to 1. No significant difference in lesion detectability among IDIFs was found, as assessed visually and using metrics such as tumor-to-background ratio (TBR) and contrast-to-noise ratio (CNR) (P < 0.05).
Conclusion:
IDIFs from smaller vessels can reliably reconstruct parametric Ki images without compromising lesion detectability, providing clinically relevant information.
More Related Videos
11:09High-Resolution Cardiac Positron Emission Tomography/Computed Tomography for Small Animals
Published on: December 16, 2022
09:21Human Brown Adipose Tissue Depots Automatically Segmented by Positron Emission Tomography/Computed Tomography and Registered Magnetic Resonance Images
Published on: February 18, 2015
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...