Tools and recommendations for commissioning and quality assurance of deformable image registration in radiotherapy
Lando S Bosma1, Mohammad Hussein2, Michael G Jameson3,4
1Department of Radiotherapy, University Medical Center Utrecht, Utrecht, the Netherlands.
Abstract:
Multiple tools are available for commissioning and quality assurance of deformable image registration (DIR), each with their own advantages and disadvantages in the context of radiotherapy. The selection of appropriate tools should depend on the DIR application with its corresponding available input, desired output, and time requirement. Discussions were hosted by the ESTRO Physics Workshop 2021 on Commissioning and Quality Assurance for DIR in Radiotherapy. A consensus was reached on what requirements are needed for commissioning and quality assurance for different applications, and what combination of tools is associated with this. For commissioning, we recommend the target registration error of manually annotated anatomical landmarks or the distance-to-agreement of manually delineated contours to evaluate alignment. These should be supplemented by the distance to discordance and/or biomechanical criteria to evaluate consistency and plausibility. Digital phantoms can be useful to evaluate DIR for dose accumulation but are currently only available for a limited range of anatomies, image modalities and types of deformations. For quality assurance of DIR for contour propagation, we recommend at least a visual inspection of the registered image and contour. For quality assurance of DIR for warping quantitative information such as dose, Hounsfield units or positron emission tomography-data, we recommend visual inspection of the registered image together with image similarity to evaluate alignment, supplemented by an inspection of the Jacobian determinant or bending energy to evaluate plausibility, and by the dose (gradient) to evaluate relevance. We acknowledge that some of these metrics are still missing in currently available commercial solutions.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET