Handheld ultrasound for portal vein evaluation in pediatric patients

Ulrich Kaiser1, Fabian Kück2, Christian Stroszczynski3

  • 1Medical Clinic and Polyclinic III, University Hospital Regensburg, Regensburg, Germany.

Insights

A handheld ultrasound device (HUD) offers sufficient image quality for pediatric portal vein (PV) assessment, though its performance decreases with deeper structures. Diagnostic capabilities are lower compared to high-end systems, particularly with increased object depth.

Area of Science:

  • Medical Imaging
  • Pediatric Radiology
  • Ultrasound Technology

Background:

  • Portal vein (PV) evaluation is crucial in pediatric diagnostics.
  • Handheld ultrasound devices (HUDs) offer portability but require validation for specific applications.
  • Assessing diagnostic capabilities of modern HUDs is essential for clinical utility.

Purpose of the Study:

  • To evaluate the diagnostic performance of a modern HUD for pediatric PV assessment.
  • To determine if patient age, body mass index (BMI), and object depth affect HUD image quality.
  • To compare HUD performance against a cart-based high-end ultrasound system (HEUS).

Main Methods:

  • 45 pediatric patients (3.9 months–17.9 years) were scanned using both HUD and HEUS.
  • B-Mode and Color-coded Doppler sonography (CCDS) were employed for PV imaging.
  • Image quality was rated on a 5-point Likert scale and correlated with patient factors and object depth (skin-to-portal-vein-distance, SPVD).

Main Results:

  • The HUD achieved image quality scores of ≥4 points (mild/no limitations) in 60% (B-Mode) and 56% (CCDS) of cases.
  • A significant negative correlation was found between HUD image quality and SPVD for both B-Mode and CCDS.
  • The influence of BMI and SPVD on image quality differed significantly between HUD and HEUS.

Conclusions:

  • HUDs provide sufficient to excellent image quality for pediatric PV assessment, except for deep-seated vessels.
  • The diagnostic performance of HUDs is lower than HEUS, primarily due to limitations with increased object depth.
  • Object depth is a key factor influencing HUD image quality in pediatric portal vein evaluations.
Abstract