Performance Evaluation of Automated Erythrocyte Sedimentation Rate (ESR) Analyzers in a Multicentric Study
Flaminia Tomassetti1,2, Cinzia Calabrese1,2, Fabio Bertani3
1Department of Experimental Medicine, University of Rome Tor Vergata, Via Montpellier 1, 00133 Rome, Italy.
Diagnostics (Basel, Switzerland)
|September 28, 2024
Summary
Three automated Erythrocyte Sedimentation Rate (ESR) analyzers demonstrated excellent accuracy and precision. These devices are reliable for clinical laboratories of all sizes, unaffected by lab environment factors.
Area of Science:
- Clinical Chemistry
- Hematology
- Laboratory Medicine
Background:
- Erythrocyte Sedimentation Rate (ESR) is a key diagnostic marker for inflammatory and infectious diseases.
- Automated analyzers offer potential for improved efficiency and standardization in ESR testing.
Purpose of the Study:
- To evaluate the performance of three automated ESR analyzers: VES-MATIC 5, CUBE 30 TOUCH, and MINI-CUBE.
- To assess inter-site variability and identify device- or operator-dependent influences across four Italian polyclinics.
Main Methods:
- Comparative analysis against the gold standard Westergren method.
- Accuracy and precision assessed using quality control materials and fresh blood samples across low, intermediate, and high ESR ranges.
- Intra-run and inter-run precision evaluated, with repeatability estimated through multiple sample re-analyses.
Main Results:
- Strong correlations observed between automated analyzers and the Westergren method (R² > 0.974).
- Excellent accuracy demonstrated with coefficients of variation (CVs) below 10% for all instruments.
- High repeatability confirmed across all ESR ranges, with CVs consistently below 10%.
Conclusions:
- All three evaluated automated ESR instruments exhibit optimal performance characteristics.
- These analyzers are suitable for diverse laboratory settings, from large to small facilities.
- Device performance is robust and not significantly influenced by laboratory environmental factors or operator variability.