Evaluation of the Impact of Infusion Set Design on the Particulate Load Induced by Vancomycin-Piperacillin/Tazobactam
Laura Négrier1, Bertrand Décaudin1, Anthony Treizebré2
1University of Lille, CHU Lille, ULR 7365-GRITA-Groupe de Recherche sur les formes Injectables et les Technologies Associées, F-59000 Lille, France.
Introduction:
Drug incompatibilities are among the most common medication errors in intensive care units. A precipitate can form and block the catheter or cause an adverse event in the patient. Intensive care units have implemented various strategies for limiting the occurrence of these incompatibilities, which have already been studied in vitro under standardized conditions. The objective of the present in vitro study was to continue these assessments by determining the impact of the infusion line geometry and the drugs' position in the infusion set-up on the prevention of vancomycin-piperacillin/tazobactam incompatibility.
Methods:
Infusion lines with a different common volume, a multilumen medical infusion device, a dilute vancomycin solution, and separate infusions of incompatible drugs were evaluated separately. The infusion line outlet was connected to a dynamic particle counter.
Results:
Reducing the common volume, using multilumen medical devices, or spacing out the two incompatible drugs on the infusion line did not prevent the occurrence of a significant particulate load. Only dilution of the vancomycin solution was associated with a significantly lower particulate load and the absence of drug incompatibility.
Conclusions:
Our results show that under specific conditions, it is possible to reduce particulate contamination considerably.
More Related Videos
09:57A Protocol to Set Up Needle-Free Connector with Positive Displacement on Central Venous Catheter in Intensive Care Unit
Published on: July 13, 2019
12:03Antimicrobial Synergy Testing by the Inkjet Printer-assisted Automated Checkerboard Array and the Manual Time-kill Method
Published on: April 18, 2019
Related Concept Videos
One-Compartment Model: IV Infusion
The one-compartment model for IV infusion uses mathematical equations to describe the rate of change in drug quantity in the body. At steady-state or infusion equilibrium, the drug input...
Endocarditis IV: Nursing Management
Two-Compartment Open Model: IV Infusion
The model illustrates the decrease in plasma drug concentration from the central compartment with a specific equation. It shows that under steady-state conditions, the drug's input rate...
Renal Failure: Dose Adjustments
Reduced renal clearance and elimination rate are common outcomes of renal impairment. These alterations lead to a prolonged elimination half-life and an altered apparent volume of distribution for drugs. As a result, dosage adjustments are typically necessary to maintain optimal drug levels in the body.
However, dosage adjustments...
Factors Influencing Drug Absorption: Pharmaceutical Parameters
Factors Affecting Renal Clearance: Drug's Physicochemical Properties and Plasma Levels
One important factor is the drug's molecular size. The kidneys readily excrete smaller molecules below 300 Daltons (Da). On the other hand, molecules weighing between 300 and 500 Da are excreted through both urine and bile. Larger molecules above 500 Da tend to be excreted...