High predicted durability for the novel small-diameter OmniaSecure defibrillation lead
George H Crossley1, Pamela K Mason2, Bert Hansky3
1Vanderbilt University Medical Center, Nashville, Tennessee.
Background:
Defibrillation leads remain the Achilles heel of implantable cardioverter-defibrillators. As patients with implantable cardioverter-defibrillators are living longer and battery longevity increases, more durable leads are needed. The LEADR trial evaluated the novel, lumenless, small-diameter, OmniaSecure defibrillation lead and demonstrated favorable safety and efficacy profile as well as zero fractures through 12.7 ± 4.8 months and remains in clinical follow-up. To augment the clinical trial, advanced cardiac lead reliability modeling was used to project long-term lead durability.
Objective:
We aimed to project the 10-year fracture-free survival of the OmniaSecure defibrillation lead using reliability modeling.
Methods:
The validated reliability model, which incorporates patient and bench test data, was used to project the 10-year fracture-free survival of the OmniaSecure lead. A subset of LEADR trial patients underwent biplane fluoroscopy imaging during cardiac and patient motion to evaluate the lead's bending curvature in vivo. Bench tests then reproduced these use conditions with greater bending curvatures than observed in patients to exaggerate stress on the lead and to evaluate the lead fatigue strength.
Results:
The reliability modeling projects a 98.2% fracture-free survival rate of the OmniaSecure lead at 10 years, including a 10-year fracture-free survival rate of 97.9% in adolescents, exceeding both the modeled and clinical 10-year performance of the highly reliable, larger diameter Sprint Quattro lead.
Conclusion:
Consistent with early clinical trial experience, modeling projects highly durable 10-year performance of the OmniaSecure lead, including within the active adolescent pediatric population, which may uniquely benefit from a novel 4.7F defibrillation lead designed for reliability.
Gov Identifier:
NCT04863664.
Related Concept Videos
Batteries and Fuel Cells
Turnover Number and Catalytic Efficiency
Chymotrypsin is a pancreatic enzyme that breaks down proteins during digestion. The...
Internal Combustion Engine
Mechanical Efficiency of Real Machines
However, in reality, no machine can be truly ideal, and all of them experience some...
Design Example: Automobile Ignition System
One can generate a large voltage using a car battery of 12 volts with the help of inductors. Inductors are known for opposing rapid...
Abrasion Resistance of Concrete
One such test is the revolving disc test, where three plates...