[Feasibility of Adapting Various Tumor-to-normal Bone Ratio Images on an Automatic Quantification Package for
Toyohiro Kato1, Hajime Ichikawa1,2, Kazunori Kawakami3
1Department of Radiology, Toyohashi Municipal Hospital.
Abstract:
We investigated the impact of the tumor-to-normal bone ratio (TNR) on the concordance rate between a detectability score classified by software (DSsoft) using an automatic quantification package for bone SPECT (Hone Graph) and a detectability score classified by visual assessment (DSvisual), and considered the feasibility of applying this software to various TNR images. 99mTc solution was filled into a SIM2 bone phantom to achieve TNRs of 4, 6, and 8, performed by dynamic SPECT acquisitions performed for 12 minutes; reconstructions were performed using ordered subset expectation maximization at timepoints ranging from 4 to 12 minutes. This yielded a total of 384 lesions (96 SPECT images). We investigated the weighted kappa (κw) coefficient between DSsoft and DSvisual at various TNRs and evaluated the change in analysis accuracy before and after applying newly created analysis parameters. DSs were defined on a 4-point scale (4: excellent, 3: adequate, 2: average, 1: poor), and visual evaluations were conducted by three board-certified nuclear medicine technologists. The κw coefficients between DSsoft and DSvisual were 0.75, 0.97, and 0.93 for TNRs 4, 6, and 8, respectively, with each κw coefficient being significant (p<0.01). In the TNR 4 image group, κw coefficients significantly increased with the implementation of new parameters proposed in this study. We concluded that the software's automatic analysis would be closer to a visual assessment within the TNR range of 4-8 and that applying new parameters derived from this study to images with TNR 4 further improves the software's automatic analysis accuracy of DSsoft. We suggest that software will be a useful tool for optimizing bone SPECT imaging techniques.
More Related Videos
08:41Patient-Specific Polyvinyl Alcohol Phantom Fabrication with Ultrasound and X-Ray Contrast for Brain Tumor Surgery Planning
Published on: July 14, 2020
11:05Tissue-simulating Phantoms for Assessing Potential Near-infrared Fluorescence Imaging Applications in Breast Cancer Surgery
Published on: September 19, 2014