Independent volumetric internal fixation reduces posterior column acetabular fracture site motion as compared to
Helyn G Fraser1, M Wesley Honeycutt2, Daniel Thompson1
1Emory University Department of Orthopaedic Surgery, USA.
Aims & Objectives:
To establish whether a suprapectineal pelvic reconstruction plate and posterior column screw (P&S) construct or a single 6.5-mm cannulated posterior column screw (PCS) construct demonstrates greater mechanical stability for fixation of acetabulum fractures involving the posterior column (PC). We hypothesized that the PCS construct would result in less fracture site motion.
Materials & Methods:
Twelve fourth-generation composite hemipelvi were utilized, 6 for each construct. The P&S construct consisted of a suprapectineal pelvic reconstruction plate with two 3.5-mm posterior column screws crossing the fracture site in lag-by-technique fashion and two screws anchoring the plate to the sciatic buttress. The PCS construct consisted of a single 6.5-mm partially threaded cannulated screw placed in an antegrade fashion. Both fixation models were cyclically loaded at 0.5 cycles/second at 400N and 800N, first in a sit-to-stand position that is expected during recovery, and subsequently in a squat-to-stand position to test overload conditions.
Results:
Under sit-to-stand loading, the PCS construct resulted in less motion at the fracture site than the P&S construct (0.06 ± 0.02 mm vs 0.1 ± 0.02 mm at 400N, p = 0.02; 0.13 ± 0.03 mm vs 0.19 ± 0.04 mm at 800N, p = 0.03). The PCS construct also demonstrated less fracture site motion under squat-to-stand loading (0.22 ± 0.13 mm vs 1.9 ± 0.5 mm at 400N, p = < 0.001; 1.48 ± 0.44 mm vs 4.77 ± 0.3 mm at 800N, p = < 0.001). At 800 N, half of the repairs failed during squat-to-stand loading (2 PCS, 4 P&S).
Conclusion:
Fixation of the posterior column of the acetabulum with a 6.5-mm cannulated screw demonstrated comparable fracture motion upon loading compared to the plate and screw construct.
More Related Videos
06:41A Minimally Invasive Model to Analyze Endochondral Fracture Healing in Mice Under Standardized Biomechanical Conditions
Published on: March 22, 2018
04:19Minimally Invasive Treatment for Thoracolumbar Burst Fracture Using Sagittal Alignment Screws and A Trauma Reduction Device
Published on: November 8, 2024
Related Concept Videos
Bone as Supporting Connective Tissue
Bone Matrix
Bone, or osseous tissue, is a connective tissue that has a large amount of two different types of matrix material. The organic matrix is similar to the matrix material found in other connective tissues, including some amount of collagen and elastic fibers. This gives strength and flexibility to the tissue. The inorganic matrix consists of mineral salts— mostly calcium salts—...
Gross Anatomy of Bone
The diaphysis is the tubular shaft that runs between the proximal and distal ends of the bone. The walls of the diaphysis are composed of dense and hard compact bone made of numerous osteons — the functional unit of the compact bone. The hollow region in the diaphysis is called the medullary cavity, which harbors the bone marrow. In infants and children, this marrow cavity is filled with red marrow, whereas in...
The Bone Matrix
Compact Bone
Compact bone, also called cortical bone, is the denser, stronger of the two types of bone tissue. It is found under the periosteum and in the diaphyses of long bones, where it provides support and protection. The microscopic structural unit of compact bone is called an osteon, or haversian system. Each osteon is composed of concentric rings of calcified...
Introduction to Joints
Spongy Bone
Spongy bone is more porous, and less dense compared to compact bone. It is composed of concentric lamellae that are arranged irregularly to form the trabecular network. In some bones, the spaces between trabeculae contain red marrow, where...