Morphological and Functional Assessment of the Right Ventricle Using 3D Echocardiography
Published on: October 28, 2020
Bedside right ventricle quantification using three-dimensional echocardiography in children with congenital heart
Khaled Hadeed1, Clément Karsenty2, Ramona Ghenghea1
1Paediatric and Congenital Cardiology, Children's Hospital, M3C CHU Toulouse, Paul Sabatier University, 31059 Toulouse, France.
Background:
Accurate quantification of right ventricular (RV) volumes and function is crucial for the management of congenital heart diseases.
Aims:
We aimed to assess the feasibility and accuracy of bedside analysis using new RV quantification software from three-dimensional transthoracic echocardiography in children with or without congenital heart disease, and to compare measurements with cardiac magnetic resonance imaging.
Methods:
We included paediatric patients with congenital heart disease (106 patients) responsible for RV volume overload and a control group (30 patients). All patients underwent three-dimensional transthoracic echocardiography using a Vivid E95 ultrasound system. RV end-diastolic and end-systolic volumes and RV ejection fraction were obtained using RV quantification software. Measurements were compared between RV quantification and cardiac magnetic resonance imaging in 27 patients.
Results:
Bedside RV quantification analysis was feasible in 133 patients (97.8%). Manual contour adjustment was necessary in 126 patients (93%). The mean time of analysis was 62±42s. RV end-diastolic and end-systolic volumes were larger in the congenital heart disease group than the control group: median 85.0 (interquartile range 29.5) mL/m2 vs 55.0 (interquartile range 20.5) mL/m2 for RV end-diastolic volume and 42.5 (interquartile range 15.3) mL/m2 vs 29.0 (interquartile range 11.8) mL/m2 for RV end-systolic volume, respectively. Good agreement for RV end-diastolic and end-systolic volumes and RV ejection fraction was found between RV quantification and magnetic resonance imaging measurements. RV quantification software underestimated RV end-diastolic volume/body surface area by 3mL/m2 and RV ejection fraction by 2.1%, and overestimated RV end-systolic volume/body surface area by 0.2mL/m2.
Conclusions:
We found good feasibility and accuracy of bedside RV quantification analysis from three-dimensional transthoracic echocardiography in children with or without congenital heart disease. RV quantification could be a reliable and non-invasive method for RV assessment in daily practice, facilitating appropriate management and follow-up care.
More Related Videos
Related Concept Videos
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion,...
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for...