Related Experiment Video
Updated: Jun 11, 2025

Untargeted Metabolomics from Biological Sources Using Ultraperformance Liquid Chromatography-High Resolution Mass Spectrometry UPLC-HRMS
Published on: May 20, 2013
Reversed-Phase Ultrahigh-Performance Liquid Chromatography-Mass Spectrometry Method for High-Throughput Lipidomic
Zuzana Vaňková1, Ondřej Peterka1, Robert Jirásko1
1Faculty of Chemical Technology, Department of Analytical Chemistry, University of Pardubice, Pardubice, Czech Republic.
Abstract:
Reversed-phase ultrahigh-performance liquid chromatography-mass spectrometry (RP-UHPLC/MS) method is optimized for the quantitation of a large number of lipid species in biological samples, primarily in human plasma and serum. The method uses a C18 bridged ethylene hybrid (BEH) column (150 × 2.1 mm; 1.7 μm) for the separation of lipids from 23 subclasses with a total run time of 25 min. Lipid species separation allows the resolution of isobaric and isomeric lipid forms. A triple quadrupole mass spectrometer is used for targeted lipidomic analysis using multiple reaction monitoring (MRM) in the positive ion mode. Data are evaluated by Skyline software, and the concentrations of analytes are determined using internal standards per each individual lipid class.
Related Concept Videos
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
High-Performance Liquid Chromatography: Elution Process