Determination of the response for the National Ignition Facility particle time of flight (PTOF) detector using single
Y Lawrence1, B L Reichelt1, C W Wink1
1Massachusetts Institute of Technology, Cambridge, Massachusetts 02139, USA.
Abstract:
The Particle Time of Flight (PTOF) detector is a chemical vapor deposition diamond-based detector used to measure bang times in low-yield (≲ 1015 neutrons) experiments at the National Ignition Facility (NIF) at Lawrence Livermore National Laboratory (LLNL). Historically, the impulse response for PTOF diamond detectors has been obtained from x-ray timing shots on the NIF and shots on the MegaRay pulsed electron accelerator at LLNL. The impulse response may alternatively be obtained using single particle interactions with the detector, at substantially less cost and higher frequency compared to NIF timing shots, which typically occur months apart. Here, the response of a PTOF detector setup is characterized by statistically averaging a large number of single particle waveforms. A high fidelity instrument response function can be constructed in this way. This is confirmed by comparison of the single particle counting-constructed response to the impulse response function measured for the same detector at LLNL's MegaRay facility.
More Related Videos
14:19A Basic Positron Emission Tomography System Constructed to Locate a Radioactive Source in a Bi-dimensional Space
Published on: February 1, 2016
11:27Studying Soft-matter and Biological Systems over a Wide Length-scale from Nanometer and Micrometer Sizes at the Small-angle Neutron Diffractometer KWS-2
Published on: December 8, 2016
Related Concept Videos
Gas Chromatography: Types of Detectors-II
Mass Analyzers: Common Types
Atomic Emission Spectroscopy: Instrumentation
Atomic Emission Spectroscopy: Lab
Double Resonance Techniques: Overview
Spin decoupling is usually achieved by...
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview