Post-irradiation darkening model for EBT-3 films characterized using a single lot calibration approach
Marco Caprioli1, Laurence Delombaerde2, Robin De Roover2
1Department of Radiation Oncology, KU Leuven, Herestraat 49, Leuven 3000, Belgium.
None:
Objective.In this study, we present a model to correct the progressive post-irradiation darkening of EBT3 films. The model allows for a clinical use of EBT3 using application and calibration films scanned with different post-irradiation times.Approach.The model is a post-irradiation time- and dose-dependent power-law function, projecting the scanned transmittance of application films to the transmittance matching the same post-irradiation time of calibration films. The model was characterized for two EBT3 production lots within the dose range 0.1-12.8 Gy. A first characterization was performed utilizing calibration films scanned repeatedly for 54 d post-irradiation (lot 1), while a fast re-characterization of a second lot used three post-irradiation times (lot 2). For a long-term validation of the model, 16 film strips were irradiated at 2 Gy on different time points starting from the day of film calibration up to 43 d afterwards (lot 1). For the multiple-dose validation of the model, 8 strips were irradiated with dose levels ranging 0-12 Gy deposited 25 d after the calibration (lot 2). As a proof of principle, the model was applied to four clinical patient-specific quality assurance film measurements with prescribed dose/fraction ranging 2.66 Gy-8 Gy.Main results. The post-irradiation transmittance decreased for higher doses up to -2.5% at 12.8 Gy, and 54 d post-irradiation. With a lot-specific model correction, the mean dose accuracy of validation strips that ranged from initial -3.4% (triple-channel) and -9.90% (blue-channel) reduced to within 3.0% (all colour channels) for doses above 1 Gy. The median dose difference with the planned dose improved from -3.5% to -1.1%, and the 3%/2 mm local gamma ranged from (48.5-92.5)% to (81.2-99.2)%.Significance.A film darkening model corrects the transmittance of EBT3 films and increases the flexibility of existing dosimetry protocols. The correction ensures dose accuracies within 3%.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
06:49In situ Grazing Incidence Small Angle X-ray Scattering on Roll-To-Roll Coating of Organic Solar Cells with Laboratory X-ray Instrumentation
Published on: March 2, 2021