Assessing the Accuracy of Fitness Smartwatch Data for Cardiovascular and Physical Activity Monitoring: A Validation Study in Digital Health
Published on: February 21, 2025
Accuracy of smartwatches for the remote assessment of exercise capacity
Alexandra Jamieson1, Siana Jones2, Nishi Chaturvedi2
1MRC Unit for Lifelong Health and Ageing, UCL, 5th Floor, 1-19 Torrington Place, London, WC1E 7HB, UK. alexandra.jamieson@ucl.ac.uk.
Abstract:
Exercise capacity is a strong independent predictor of cardiovascular and all-cause mortality. The utilization of well-established submaximal tests of exercise capacity such as the 6-min walk test (6MWT), 3-min step test (3MST) and 10-chair rise test (10CRT) in the community would improve patient care but requires remote monitoring technology. Consumer grade smartwatches provide such an opportunity, however, their accuracy in measuring physiological responses to these tests is unclear. The aim of this study was to determine the accuracy of consumer grade smartwatches in assessing exercise capacity to develop a framework for remote, unsupervised testing. 16 healthy adults (7 male (44%), age median 27 [interquartile range (IQR) 26,29] years) performed 6MWTs using two protocols: (1) standard-straight 30 m laps (6MWT-standard) and 2) continuous lap-circular 240 m laps around a park (6MWT-continuous lap), 3MSTs and 10CRTs. Each one of these four tests was performed three times across two clinic visits. Each participant was fitted with a Garmin Vivoactive4 and Fitbit Sense smartwatch to measure three parameters: distance, step counts and heart rate (HR) response. Reference measures were a meter-wheel, hand tally counter and ECG, respectively. Mean HR was measured at rest, peak exercise and recovery. Agreement was measured using Bland-Altman analysis for repeated measures and summarized as median absolute percentage errors (MAPE). Distance during 6MWT-continuous lap had better agreement than during 6MWT-standard for both Garmin (MAPE: 6.4% [3.0, 10.4%] versus 20.1% [13.9, 28.4%], p < 0.001) and Fitbit (8.0% [2.9, 10.1% versus 18.8% [15.2, 28.1%], p < 0.001). Garmin measured step count more accurately than Fitbit (MAPE: 1.8% [0.9, 2.9%] versus 8.0% [2.6, 12.3%], p < 0.001). Irrespective of test, both devices showed excellent accuracy in measuring HR at rest and recovery (≤ 3%), while accuracy decreased during peak exercise (Fitbit: ~ 12% and Garmin: ~ 7%). In young adults without mobility difficulties, exercise capacity can be measured remotely using standardized tests and consumer grade smartwatches.
Related Concept Videos
Exercise Stress Test
Exercise stress testing, commonly known as a treadmill test, is a noninvasive procedure used to evaluate cardiovascular function and diagnose heart conditions.
Definition
An exercise stress test measures the heart's response to exertion using a treadmill or stationary bicycle. Chest electrodes record the heart's electrical activity through an ECG, and blood pressure is monitored regularly.
Purposes
Holter Monitor: 24-Hour Monitoring
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)