Related Experiment Video
Updated: Jun 11, 2025

Full-root Aortic Valve Replacement by Stentless Aortic Xenografts in Patients with Small Aortic Roots
Published on: May 21, 2017
Pannus formation: a rare culprit of early bioprosthetic valve dysfunction-a case report
Sho Takemoto1,2, Hiroshi Kumano3, Junichi Shimamura4
1Center for Transplantation Sciences, Department of Surgery, Massachusetts General Hospital/Harvard Medical School, 13th Street, Building 149, Boston, MA 02129, USA.
Background:
Early bioprosthetic valve dysfunction (BVD) due to pannus formation is uncommon in elderly patients, and only a limited number of cases have been reported.
Case Summary:
An 84-year-old man presented with exertional dyspnoea 3 years after surgical aortic valve replacement (SAVR) with a 19 mm Epic™ valve (Abbott, Santa Clara, CA, USA). Transthoracic echocardiography demonstrated progressive BVD, and cardiac computed tomography (CT) revealed sub-aortic pannus formation. Re-operative SAVR was performed using a 19 mm INSPIRIS RESILIA® valve (Edwards Lifesciences, Irvine, CA, USA), and pathological examination confirmed valve leaflet deformation caused by pannus overgrowth. At the 18-month follow-up, the patient exhibited favourable progress, with no indications of BVD or pannus recurrence.
Discussion:
This case highlights the importance of recognizing early pannus formation as a cause of BVD, even in elderly patients. Early detection of BVD based on clinical symptoms and echocardiography is vital to allow timely surgical intervention before the deterioration of cardiac function. Cardiac CT helps to differentiate pannus from thrombus formation and guide treatment decisions.