Related Experiment Video
Updated: Jun 11, 2025

Evaluation of the Efficacy of Organic Peroxyacids for Eradicating Dairy Biofilms Using an Approach Combining Static and Dynamic Methods
Published on: December 9, 2022
Effect of simethicone on the bactericidal efficacy of a high-level disinfectant
Gregory G Anderson1, Katharine Segars2, Anastacia M Sanchez1
1Division of Biology, Chemistry, and Materials Science, Office of Science and Engineering Laboratories, Center for Devices and Radiological Health, US Food and Drug Administration, Silver Spring, MD 20993, USA.
Abstract:
Introduction. Simethicone is an over-the-counter product that is frequently used by clinicians during endoscopic procedures to reduce foaming and improve visualization. Published studies have found simethicone residue on endoscopes after cleaning and disinfecting the devices as per the manufacturer's instructions. Some literature suggests that simethicone residue may reduce disinfection efficacy and increase the risk of patient infections.Gap Statement. However, there appears to be a lack of direct evidence in the literature to either disprove this or correlate simethicone presence with an increased microbial risk.Aim: Research was conducted to evaluate the in vitro impact of simethicone on disinfection efficacy.Methodology. Bacteria were grown in a microtitre plate assay in the presence of a range of simethicone concentrations and then treated with a disinfectant. Bacterial growth was assessed by spotting each microtitre well onto an agar plate.Results. The results demonstrated that, under the conditions tested, simethicone did not reduce the efficacy of Cidex ortho-phthalaldehyde disinfectant, which demonstrated at least a 6-log unit reduction in bacterial viability. Additional experiments showed that direct exposure to 66 mg ml-1 of simethicone reduced bacterial viability.Conclusion. These results indicate that simethicone may not reduce the bactericidal efficacy of disinfectant during reprocessing, under certain conditions.
Insights
Simethicone, used during endoscopies, does not appear to reduce the effectiveness of Cidex ortho-phthalaldehyde disinfectant. This study found no increased microbial risk from simethicone residue on reprocessed endoscopes.
Area of Science:
- Microbiology
- Infection Control
- Medical Device Reprocessing
Background:
- Simethicone is commonly used in endoscopic procedures to improve visualization.
- Concerns exist that simethicone residue on endoscopes may compromise disinfection efficacy and increase infection risk.
- Limited direct evidence addresses the impact of simethicone on microbial risk during reprocessing.
Purpose of the Study:
- To evaluate the in vitro impact of simethicone on the efficacy of disinfectants.
- To determine if simethicone residue affects the bactericidal activity of disinfectants used for endoscopes.
Main Methods:
- Bacteria were cultured in microtitre plates with varying simethicone concentrations.
- Bacterial cultures were subsequently treated with a disinfectant (Cidex ortho-phthalaldehyde).
- Bacterial viability was assessed by agar plating to determine disinfectant efficacy.
Main Results:
- Simethicone did not reduce the efficacy of Cidex ortho-phthalaldehyde disinfectant.
- The disinfectant achieved at least a 6-log unit reduction in bacterial viability.
- High concentrations of simethicone (66 mg/ml) demonstrated direct bactericidal effects.
Conclusions:
- Simethicone does not appear to impair the bactericidal efficacy of ortho-phthalaldehyde disinfectant under tested conditions.
- The presence of simethicone residue may not increase microbial risk during endoscope reprocessing.
- Further research may be warranted to explore simethicone's interaction with other disinfectants and conditions.
Related Concept Videos
Antimicrobial Effectiveness
Chemical Agents for Microbial Control
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Hand hygiene
Hand washing...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Physical Methods for Controlling Microbial Growth: Radiation and Filtration