Improved metal suppression using new generation low-field MRI: a biophantom feasibility study
Johanna Luitjens1,2, Katharina Ziegeler3, Daehyun Yoon1
1Department of Radiology and Biomedical Imaging, University of California San Francisco, 185 Berry Street, San Francisco, CA, 94158, USA.
Objective:
Novel 0.55 MRI scanners have the potential to reduce metal artifacts around orthopedic implants. The purpose of this study was to compare metal artifact size and depiction of anatomy between 0.55 T and 3.0 T MRI in a biophantom.
Materials And Methods:
Steel and titanium screws were implanted in 12 porcine knee specimens and imaging at 0.55 T and 3 T MRI was performed using the following sequences: turbo spin-echo (TSE), TSE with view angle tilting (VAT), and slice encoding for metal artifact correction (SEMAC) with proton-density (PD) and T2-weighted short-tau inversion-recovery (T2w-STIR) contrasts. Artifacts were measured, and visualization of anatomy (cartilage, bone, growth plates, cruciate ligaments) was assessed and compared between groups.
Results:
Metal artifacts were significantly smaller at 0.55 T. The smallest artifact sizes were achieved with SEMAC at 0.55 T for both PD and T2w-STIR sequences; corresponding relative size reductions vs. 3.0 T were 78.7% and 79.4% (stainless steel) and 45.3% and 1.4% (titanium). Depiction of anatomical structures was superior at 0.55 T.
Conclusion:
Substantial reduction of artifact size resulting in superior depiction of anatomical structures is possible on novel 0.55 T MRI systems. Further clinical studies are required to elucidate patient-relevant advantages.
More Related Videos
17:16Registered Bioimaging of Nanomaterials for Diagnostic and Therapeutic Monitoring
Published on: December 9, 2010
08:19Protocol for the Evaluation of MRI Artifacts Caused by Metal Implants to Assess the Suitability of Implants and the Vulnerability of Pulse Sequences
Published on: May 17, 2018
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System IV: CMRI