Evaluating the accuracy and reliability of non-piezo portable ultrasound devices in postpartum care
Ruben Plöger1, Charlotte Behning2, Adeline Walter1
1Department of Obstetrics and Prenatal Medicine, University Hospital Bonn, Venusberg Campus 1, 53127, Bonn, Germany.
Introduction:
The early diagnosis of hemorrhage via postpartum ultrasound is crucial to initiate therapy and, thus, prevent maternal death. In these critical situations rapid availability and simple transport of ultrasound devices is vital, paving the way for a new generation of portable handheld ultrasound devices (PUD) consisting of transducers and tablets or smart phones. However, evidence to confirm the diagnostic accuracy of these new devices is still scarce.
Methods:
The accuracy and reliability of these new devices in relation to established standard ultrasound devices is analyses in this pilot study by comparing diagnoses and by applying statistical analysis via Bland-Altman plots, intraclass correlation coefficients (ICC), and Pearson correlation coefficients (PCC). One hundred patients of a university hospital were included in this study.
Results:
In all cases, the same diagnosis was made regardless of the applied ultrasound device, confirming high accuracy. There was a high correlation (PCC 0.951) and excellent agreement (ICC 0.974) in the assessment of the cavum, while the assessment of the diameters of the uterus showed only a good correlation and a good agreement. Subgroup analysis for maternal weight, mode of delivery and day after delivery was performed CONCLUSION: The same diagnosis independent of the used devices and excellent results of the cavum assessment promote the use of PUDs in a clinical setting. The slightly lower accuracy in the measurement of the uterus may be caused by the PUD's small acoustic window, reflecting one of its weaknesses. Therefore, the patient may benefit from the short time to diagnosis and the unbound location of examination, either in the delivery room, on the ward, or at home.
More Related Videos
11:27A Modified Sonographic Algorithm for Image Acquisition in Life-Threatening Emergencies in the Critically Ill Newborn
Published on: April 7, 2023
06:15Protocol and Guidelines for Point-of-Care Lung Ultrasound in Diagnosing Neonatal Pulmonary Diseases Based on International Expert Consensus
Published on: March 6, 2019
Related Concept Videos
Ultrasound I: Abdominal Ultrasonography
Abdominal ultrasonography, commonly known as abdominal ultrasound, is a vital, non-invasive medical imaging technique widely used in healthcare.
Procedure:
This diagnostic tool allows the clinician to visually inspect internal structures within the abdomen, including vital organs such as the liver, gallbladder, pancreas, kidneys, and spleen.
The abdominal ultrasound process begins with applying a special gel to the patient's skin over the abdomen. This gel enhances the...
Ultrasonography
During an ultrasonography procedure, a handheld device called...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: