Accuracy of robotic arm-assisted versus computed tomography-based navigation in total hip arthroplasty using the
Tomoya Okazaki1,2, Takashi Imagama3, Yuta Matsuki3,4
1Department of Orthopedic Surgery, Yamaguchi University Graduate School of Medicine, Ube, Yamaguchi, 755-8505, Japan. tomoyaokazaki0407@gmail.com.
Background:
A robotic arm-assisted and a computed tomography (CT)- based navigation system have been reported to improve the accuracy of component positioning in total hip arthroplasty (THA). However, no study has compared robotic arm-assisted THA (rTHA) to CT-based navigated THA (nTHA) concerning accuracy of cup placement and acetabular fractures using the direct anterior approach (DAA). This study aimed to compare the accuracy of cup placement and the presence of intraoperative acetabular fractures between rTHA and nTHA using DAA in the supine position.
Methods:
We retrospectively investigated 209 hips of 188 patients who underwent rTHA or nTHA using DAA (rTHA using the Mako system: 85 hips of 79 patients; nTHA: 124 hips of 109 patients). After propensity score matching for age and sex, each group consisted of 73 hips. We evaluated clinical and radiographic outcomes, comparing postoperative cup orientation and position, measured using a three-dimensional templating software, to preoperative CT planning. Additionally, we investigated the prevalence of occult acetabular fracture.
Results:
Clinical outcomes were not significantly different between the groups at 1 year postoperatively. The mean absolute error of cup orientation was significantly smaller in the rTHA group than in nTHA (inclination: 1.4° ± 1.2° vs. 2.7° ± 2.2°, respectively; p = 0.0001, anteversion: 1.5° ± 1.3° vs. 2.2° ± 1.7°, respectively; p = 0.007). The cases within an absolute error of 5 degrees in both RI and RA were significantly higher in the rTHA (97.3%) than in nTHA group (82.2%) (p = 0.003). The absolute error of the cup position was not significantly different between the two groups. The prevalence of occult acetabular fracture did not differ significantly between the two groups (rTHA: n = 0 [0%] vs. nTHA: n = 1 [1.4%]).
Conclusion:
Cup placement using DAA in the supine position in rTHA was more accurate with fewer outliers compared to nTHA. Therefore, rTHA performed via DAA in a supine position would be useful for accurate cup placement.