Effective doses of scout projections in maxillofacial cone beam computed tomography
Matheus Barros-Costa1, Jack Britten2, Colin Jones2
1Department of Oral Diagnosis, Piracicaba Dental School, State University of Campinas, Piracicaba, SP, Brazil.
Objectives:
To assess the effective and organ/tissue equivalent radiation doses of different scout projection protocols in four CBCT units.
Methods:
Optically stimulated luminescence dosimeters (OSLD) were placed in reference anatomical locations in the head and neck segments of an anthropomorphic phantom representing an average adult male. Ten repeated exposures were obtained from each of the twelve scout projections studied, acquired from four maxillofacial cone beam computed tomography (CBCT) units (Midmark EIOS, 3D Accuitomo F170, Veraviewepocs 3D R100, and Veraview X800). The effective and organ/tissue equivalent doses were calculated for each protocol.
Results:
Effective doses ranged from 0.7 µSv (Accuitomo F170 60 × 60 mm-anterior maxilla) to 6.9 µSv (Midmark 50 × 50 mm-anterior maxilla). The highest organ/tissue equivalent doses were recorded for the oral mucosa and salivary glands; however, the thyroid was the highest contributor to the effective dose, followed by the salivary glands.
Conclusions:
Despite some variability among CBCT machines and protocols, the acquisition of scout projections is a low-dose procedure. The use of scout projections to ensure an adequate position of the region of interest within the field of view is highly desirable as they contribute minimally to overall CBCT patient dose.