Effect of different patient positions on endotracheal tube cuff pressure in patients undergoing urological
Eda Tok1, Nursen Karaca1, Ozge Karakoc1
1From the Department of Anesthesiology and Reanimation, Ege University School of Medicine, Izmir, Turkey.
Background:
The endotracheal tube (ETT) contains a cuff that is placed in the trachea to prevent gas leakage and aspiration of secretions and gastric contents. However, patient positioning after intubation may cause ETT displacement and changes in cuff pressure.
Objectives:
Evaluate the effect of different patient positions on ETT cuff pressure in patients undergoing urological procedures in supine, prone, lateral flank, and lithotomy positions.
Design:
Prospective and observational study.
Setting:
A university hospital in Turkey.
Patients And Methods:
Patients who underwent surgeries under general anesthesia in different patient positions were involved. After intubation (T0), the cuff pressure was checked with a manometer and adjusted to 25 cmH2O and continuously monitored. The cuff pressure was checked before (T1) and after achieving the final position (T2) and then at 5, (T3), 10, (T4), 15 minutes (T5) of the position, at the end of the procedure (T6) and before extubation (T7). At postoperative 2nd and 12th hours, the patients were interviewed for sore throat, hoarseness, and cough.
Main Outcome Measures:
The effect of different patient positions on the ETT cuff pressure.
Sample Si̇Ze:
200 patients.
Results:
The cuff pressure increased significantly at T2 in the lithotomy, lateral flank, and prone groups (P<.001 each). The highest increase in cuff pressure occurred in the prone group (34.3 [7.5] cmH2O). Over time, the cuff pressure decreased in all groups during surgery. Postoperative complications at the 2nd postoperative hour were similar in all groups; however, the mean cuff pressure was significantly higher in the patients with postoperative sore throat or cough (sore throat: P=.003; cough: P=.047).
Conclusion:
ETT cuff pressures are affected by different patient positioning; therefore, regular recording and adjustment of cuff pressure are necessary for patient safety.
Limitation:
We used ETT of a single manufacturer. Therefore, our findings may not be applicable to other types of ETT.
Related Concept Videos
Pre-Procedural Guidelines for Assessing Blood Pressure
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Endoscopic Studies II: Thoracocentesis
Description
Excess pleural fluid or air may accumulate in some respiratory disorders in the thoracic cavity. To treat pleural effusion, a physician conducts thoracentesis by carefully piercing the chest wall and entering...