Related Experiment Video
Updated: Jun 11, 2025

Techniques of Laparoscopic Right Posterior Sectionectomy: Glissonian Approach and a Parenchymal Transection Technique
Published on: October 6, 2023
Adverse events of the endoscopic over-the-scope clips and cutters: a Manufacturer and User Facility Device Experience
Ahmed Telbany1, Islam Mohamed2, Khaled Alsabbagh Alchirazi3
1Department of Gastroenterology & Hepatology, University of New Mexico, Albuquerque, New Mexico, USA.
Background And Aims:
The Ovesco clip system (Tübingen, Germany), including the over-the-scope clip and full-thickness resection device, has demonstrated efficacy in managing various GI pathologies. However, real-world data on device malfunctions and patient adverse events are limited. This study aimed to analyze adverse events associated with Ovesco clips and cutters reported to the U.S. Food and Drug Administration Manufacturer and User Facility Device Experience (MAUDE) database.
Methods:
The MAUDE database was searched for reports related to Ovesco clips and cutters from January 1, 2007, to August 30, 2024. Event date, device type and model, manufacturer, event type, device problem, and patient adverse events were extracted. Descriptive statistics were used to summarize the findings.
Results:
Forty-two reports were identified, with 41 involving Ovesco clips and 1 involving an Ovesco cutter. The most common device problem for Ovesco clips was failure to deploy (61.0%), followed by unintended deployment or misfiring (17.1%). GI perforation was the most frequently reported patient adverse event (60%). The Ovesco cutter report described device component breakage and foreign body retention. No clear trends in adverse event reporting were observed over the study period.
Conclusions:
Ovesco clips and cutters are associated with potential device malfunctions and patient adverse events, particularly GI perforation. The findings highlight the need for careful patient selection, meticulous technique, and close postprocedural monitoring. Collaborative efforts among stakeholders are essential to optimize device safety and efficacy. Continued postmarketing surveillance and real-world data analysis are crucial for monitoring the performance of endoscopic devices and improving patient outcomes.
Related Concept Videos
Endoscopic Procedures I: Esophagogastroduodenoscopy
During an EGD, the endoscope can be used to:
Endoscopic Procedures V: ERCP
Patient...
Endoscopic Procedures IV: Sigmoidoscopy and Laproscopy
Sigmoidoscopy
Sigmoidoscopy is a diagnostic procedure that uses a flexible sigmoidoscope equipped with a light source and camera to examine the rectum and sigmoid colon. The procedure involves inserting the tube through the anus...
Endoscopic Procedures III: Video Capsule Endoscopy
Endoscopic Studies II: Thoracocentesis
Description
Excess pleural fluid or air may accumulate in some respiratory disorders in the thoracic cavity. To treat pleural effusion, a physician conducts thoracentesis by carefully piercing the chest wall and entering...
Esophageal Perforation-II: Clinical Manifestations and Management
Clinical Manifestations: