Continuous Venous-Arterial Doppler Ultrasound During a Preload Challenge
Published on: January 20, 2023
Small changes in the transducer position cause a systematic change in cardiac output readings: implications for
Caroline Dinesen1,2, Simon Tilma Vistisen2, Peter Juhl-Olsen3,4
1Department of Cardiothoracic- and Vascular Surgery, Aarhus University Hospital, Anesthesia Section, Palle Juul-Jensens Boulevard 99, Aarhus N, 8200, Denmark.
None:
To systematically evaluate the effect of small changes in transducer position on key hemodynamic variables including CO generated by 4th generation FloTrac software. After cardiac surgery, cardiac output, mean arterial pressure, systemic vascular resistance, and stroke volume variation were measured with 4 generation Flotrac software. The transducer position was randomly placed at the midaxillary plane, 4 cm higher than the midaxillary plane or 4 cm lower than the midaxillary plane. Averages of three measurements were used. Data was available from 20 patients. Cardiac output increased from 4.59 L/min (± 0.92) to 4.78 L/min (± 0.99) with the transducer position at the midaxillary plane to 4 cm higher than the midaxillary plane, and cardiac output decreased to 4.43 L/min (± 0.90) with the transducer 4 cm lower than midaxillary plane (P < 0.001). On the relative scale, CO increased 4.1% (95% CI 3.1-5.0) when comparing the higher transducer level with the midaxillary plane position, and CO decreased 3.4% (95% CI 2.4-4.4) when comparing the midaxillary plane position with the lower transducer level, correspondiong to changes in CO of ≈ 1% per 1 cm change in transducer position. Mean arterial pressure and systemic vascular resistance both changed significantly with transducer position (both P < 0.001), whereas no statistically or clinically significant effect was seen on stroke volume variation (P = 0.98). A four-centimeter change in vertical transducer position induced clinically significant changes in cardiac output measurements by 4th generation FloTrac software. Definitions of optimal cardiac output in goal-directed therapy algorithms require meticulous transducer adjustment and can only be used in the reference patient position.
Related Concept Videos
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Pre-Procedural Guidelines for Assessing Blood Pressure
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion,...
Pulse
Pulse Rate and its Significance
Pulse rate, often measured in beats per minute (bpm), reflects the heart rate (HR), which is influenced by numerous factors such as stress, physical activity, and hormonal changes. A normal resting adult pulse rate falls...