Personalization of pharmacotherapy with sirolimus based on volumetric absorptive microsampling (VAMS) in pediatric
Arkadiusz Kocur1,2, Agnieszka Czajkowska3, Kamila Rębis4
1Department of Drug Chemistry, Pharmaceutical and Biomedical Analysis, Faculty of Pharmacy, Medical University of Warsaw, Banacha 1, Warsaw, 02-097, Poland. arkadiusz.kocur@wum.edu.pl.
Background:
The benefits of pharmacotherapy with sirolimus (SIR) in pediatric transplant recipients are well established. Traditionally, whole blood samples have been used to measure SIR concentrations. Volumetric Absorptive Microsampling (VAMS) is an alternative sampling strategy suitable for Therapeutic Drug Monitoring (TDM). In this study, we developed and validated two liquid chromatography-tandem mass spectrometry (LC-MS/MS) methods for determining SIR concentrations in whole blood (WB) and capillary whole blood samples collected using a VAMS-Mitra™ device.
Methods:
We used protein precipitation during WB sample preparation and dispersive liquid-liquid microextraction (DLLME) with methyl tert-butyl ether for VAMS sample preparation to optimise the analyte extraction process. The described validation protocols were cross-validated, confirming the equivalence of the whole-blood and VAMS-based methods. Furthermore, the developed methods were evaluated in two three-level rounds of an external proficiency-testing scheme.
Results:
The analytical methods were successfully validated within the calibration range of SIR (0.5-60 ng/ml). The validation parameters met the European Medicines Agency (EMA) and the International Association of Therapeutic Drug Monitoring and Clinical Toxicology (IATDM&CT) acceptance criteria. No hematocrit (tested in the range of 24.3-64.1%), matrix, or carry-over effects were observed. Cross-validation confirmed the interchangeability between VAMS-LC-MS/MS and WB-LC-MS/MS methods. The developed methods were successfully implemented for SIR determination in 140 clinical samples (70 each of WB and VAMS) from pediatric renal transplant recipients, demonstrating their practicality and reliability.
Conclusion:
The VAMS-based method has been rigorously tested and is clinically equivalent to the reference WB-LC-MS/MS method. Additionally, clinical validation confirmed the utility of the presented methods for TDM of the SIR in the pediatric population after renal transplantation.